Share on:
Model | Bante930 |
pH | |
Range | -2.000~20.000pH |
Accuracy | ±0.002pH |
Resolution | 0.1, 0.01, 0.001pH |
Calibration Points | 1 to 5 points |
Calibration Solutions | USA, NIST, DIN or User-defined |
mV | |
Range | -1999.9~1999.9mV |
Accuracy | ±0.2mV |
Resolution | 0.1, 1mV |
Calibration Points | 1 point (Only for Relative mV mode) |
Calibration Range | ±200mV |
Ion | |
Range | 0.001~19999ppm, mg/L, mol/L (Refer to the range of ISE) |
Accuracy | ±0.5% F.S (Monovalent), ±1% F.S (Divalent) |
Resolution | 0.001, 0.01, 0.1, 1 |
Calibration Points | 2 to 5 points |
Calibration Solutions | 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mg/L, mol/L |
Temperature | |
Range | 0~105°C, 32~221°F |
Accuracy | ±0.5°C, ±0.9°F |
Resolution | 0.1°C |
Calibration Points | 1 point |
Calibration Range | Measured value ±10°C |
Others | |
Temperature Compensation | 0~100°C, 32~212°F |
Compensation Mode | Manual or Automatic |
Hold Function | Manual or Automatic |
Stability Conditions | Low or High |
Calibration Due | 0 to 31 days |
Power Off | Manual or Automatic (10, 20, 30 minutes) |
Reset Function | Yes |
Memory | Stores up to 500 data sets |
Output | USB Communication Interface |
Connector | BNC |
Display | LCD (130×110mm) |
Power Requirements | DC5V, using AC adapter, 220VAC/50Hz |
Dimensions | 210(L)×188(W)×60(H)mm |
Weight | 1.5kg |
Order Code | Dimensions | Weight |
Bante930-CN | 310(L)×240(W)×180(H)mm | 1.7kg |
Bante930-UK | 310(L)×240(W)×180(H)mm | 1.7kg |
Nicole Chen | ||||||
TEL: | +86-571-87571061 | WECHAT: | 2856943842 | |||
EMAIL: | sales2@westtune.com | WHATSAPP: | +8613184235418 | |||
WEB: | www.westtune.com | SKYPE: | nicolewesttune@qq.com | |||
ADD: | Room 502, Building B, No.359, Shuxin Road, Hangzhou, Zhejiang, China |