Share on:
We have -20-150, -30-150, 50-650, 300-1200, 100-1200degree, if you need, please contact me to send the corresponding instructions to you ,thanks
1. Use profile:
Used in the required temperature range, it is easy to set and accurately produce the required temperature.It has the characteristic of small volume, fast heating, stable temperature control, no pollution, etc.The temperature sensor and instrument can be integrally calibrated, so it is easy to carry, easy to use, easy to fast and reliable temperature calibration. It is widely used in machinery, chemical, food, pharmaceutical and other industries.
2. Product features:
2. 1. Various types of insert tubes can meet the test and calibration of sensors of different sizes and quantities.
2. 2. The heating rate can be adjusted according to user requirements
temperature range | -20℃~150℃ | -30℃~150℃ | 50℃~650℃ | 300℃~1200℃ |
heating rate
| 30~100:20min 30~150:40min | 30~100:20min 30~150:40min | 30~300:7min 30-400:12min 30~650:25min | 30~1200:75min |
Display accuracy | ≤±0.1℃ | ≤±0.1℃ | Below 400℃:≤±0.35℃ 400-650℃:≤±0.5℃ | ≤±1.2℃ |
Insertion depth | 160mm | 150mm | 135mm | |
Average heat block diameter | 36mm | 32mm | 32mm | 39mm |
Temperature field stability | ≤±0.02℃ | ≤±0.02℃ | ≤±0.05℃ | ≤±0.2℃ |
Horizontal temperature field | ≤±0.05℃ | ≤±0.05℃ | ≤±0.05℃ | ≤±0.25℃ |
Vertical temperature field | The deviation in the range of 50mm calculated from the bottom of the hole of the soaking block is 1 degree | |||
Temperature unit | ℃ or ℉ | |||
Display resolution | 0.01℃ or 0.1℃ | 0.01℃ | 0.01℃ or 0.1℃ | 0.01℃ or 0.1℃ |
maximum power | Negative temperature type: 500W, medium temperature type: 800W, high temperature type: 3000W | |||
weight(Net weight) | Negative temperature type: 13kg, medium temperature type: 11kg, high temperature type: 11kg | |||
weight (Containing packaging)
| Negative temperature type: 23kg, medium temperature type: 18kg, high temperature type:20kg Packaging includes aluminium box and transport wooden box | |||
Outline size
| Negative temperature type and high temperature type: 310*190*340mm, medium temperature type: 250*150*310mm | |||
Use environment | Ambient temperature 0-50℃, relative humidity less than 95% (no condensation) | |||
Power supply | 220VAC ± 10%, 45 ~ 60Hz, optional 110VAC ± 10%. | |||
Selection function | Verification Data Recording and Exporting (ET2501 only optional), up to 250 sensors, 2500 records,USB Device Communication Interface |