Technical specifications:
* AC Power supply: AC220 V-230 V/50 Hz or AC110 V-120 V/50-60 Hz
* DC Power supply: DC12 V/10 A or DC24 V/6 A
* Input Power: 400 VA
* Mode of control: Air mode and baby mode are controlled by micro-computer
* Air temperature control range: 25°C -37°C (Override mode 37-38°C)
* Baby temperature control range: 34°C -37°C (Override mode 37-37.5°C)
* Skin temperature sensor accuracy: ± 0.3°C
* Temperature variability: ≤ 1.0°C
* Temperature uniformity: ≤ 1.5°C (mattress is on the horizontal position)
* Noise inside hood: ≤ 52 dB(A)
* Failure alarm: over temperature alarm, temperature deviation alarm, sensor alarm, Motor
* Fan Failed Alarm, Power failure Alarm and so on
* Operating at temperature equilibrium: 90 min (Set Temp 36°C - Ambient 15°C)
Standard configuration:
* Main body (including the hood, controller bassinet, internal battery and observation lamp),
oxygen cylinders, oxygen supply system, skin temperature sensor, I.V. Pole, mattress,
adjustable stand.
*
Optional configuration:
* > 37°C override temperature mode;
* Disposable skin temperature sensor;
* Ambulance trolley.
Environmental Temperature:
Operating range: 18°C - 30°C
Ambient air movement rate: ≤ 0.3 m/s
Product Packaging:
Packing: each unit is packed into one carton
Dimensions: 127 x 82 x 113.5 cm
Gross Weight: 100 kg (HKN-93), 110 kg (HKN-93B), 132 kg (HKN-93A).