Item | Value |
Type | Body Fluids Processing Equipments |
Brand Name | Fresenius |
Place of Origin | Germany |
Instrument classification | Class I |
Warranty | 2 years |
After-sale Service | Online technical support |
Dimensions | 1370x500x650(HxWxD) |
Weight | 86 KG |
Water supply Water inlet pressure | 1.5 to 6.0 bar |
Water supply Water inlet temperature | 5°C to 30°C; for “integrated hot rinse” 85°C to 95°C |
Water supply Max. drain height | 1 m |
Concentrate Supply pressure | 0 to 100 mbar; 1 m max. suction height |
Concentrate Central supply | 1 central acid concentrate (optional) 0 to 500 mbar |
Power supply | 100 to 240 V AC, 50 to 60 Hz |
Current consumption | Approx. 9 A (at 230 V) and 15 A (at 110 V) |
External connections | Network interface for data exchange with Therapy Data Management System (optional); input/output for connection of external auxiliary equipment; alarm input/output (e.g. for nurse call); diagnosis for in-house computer diagnosis |
Battery | 18 V; 3 Ah (maintenance free) |
Arterial pressure monitoring Display range | - 300 mmHg to + 280 mmHg |
Arterial pressure monitoring Accuracy | ±10 mmHg |
Arterial pressure monitoring Resolution | 20 mmHg |
Venous pressure monitoring Display range | - 60 mmHg to + 520 mmHg |
Venous pressure monitoring Accuracy | ±10 mmHg |
Venous pressure monitoring Resolution | 20 mmHg |
Transmembrane pressure monitoring Display range | - 60 mmHg to + 520 mmHg |
Transmembrane pressure monitoring Resolution | 20 mmHg |
Arterial blood pump Blood flow range | 15 to 600 mL/min |
Arterial blood pump Accuracy | ± 10% |
Arterial blood pump Resolution | 5 mL/min |
Arterial blood pump Single needle system (optional) | With 2 blood pumps, internal pressure/pressure control with variable stroke volume (max. 50 mL/min) |
Arterial blood pump Air bubble detector | Ultrasonic transmission measurement on blood line, additional optical monitoring in venous clamp |
Arterial blood pump Heparin pump | Delivery range: 0 to 10 mL/h Bolus function: max. 5 mL per bolus Syringe size: 20 mL |
Dialysis fluid flow range Dialysis fluid flow range | 0 – 300 – 500 – 800 mL/min |
Dialysis fluid flow range AdaptedFlow*(selectable) | Dialysate flow adapted to the effective blood flow |
Dialysis fluid temperature Selectable | 35°C to 39°C |
Dialysis fluid conductivity Range | 12.8 to 15.7 mS/cm (25°C) |
Dialysis fluid conductivity Accuracy | ± 0.1 mS/cm |
Dialysis fluid conductivity Resolution | 0.1 mS/cm |
Dialysis fluid acid component Mixing ratio | Adjustable, e.g. 1+44, 1+34 |
Dialysis fluid acid component Adjustment range | 125 to 150 mmol/L |
Dialysis fluid bicarbonate component Default mixing ratio | 1 + 27.6 (others possible) |
Dialysis fluid bicarbonate component Adjustment range | - 8 to + 8 mmol/L |
Accurate Clearance K | ± 6% |
Bicarbonate dry concentrate | bibag |
Dialysis fluid filter system | DIASAFEplus |
Balancing accuracy | ± 0.1 % according to the total dialysate volume |
Pressure holding tests | Cyclic |
Ultrafiltration UF rate | 0 to 4000 mL/h (in steps of 1 mL/h) |
Ultrafiltration Pump volume accuracy | ± 1% |
Ultrafiltration Parameters displayed | UF goal, UF time, UF rate, UF volume |
Blood leak detector Sensitivity | ≤ 0.5 mL blood/min (Hct = 25) at max. flow 800 mL/min |
Disinfection and cleaning programmes** | |
Rinse Temperature/flow | 37 °C / 600 mL/min |
Hot rinse (recirculation) Temperature/flow | 84 °C / 450 mL/min |
Integrated hot rinse Temperature/flow | 84°C / 450 mL/min |
Cleaning Sporotal100 (recirculation) Temperature/flow | 37 °C / 600 mL/min |
Heat disinfection Diasteril/Citrosteril (recirculation) Temperature/flow | 84° C / 450 mL/min |
Disinfection Puristeril 340/plus (recirculation) Temperature/flow | 37 °C / 600 mL/min |