Share on (1600353174403):



MODELS | EA313 |
infusion rate (ml/h) | 1~400 ml/hr in 1 ml increments |
mean flow velocity accuracy | ±10% of selected flow rate |
dosage limit(ml) | 0-9999 in 1 ml increment |
meanvolume accuracy | ±10% |
loading speed andmeanaccuracy (ml/h) | 1000±10% |
Cleaning speed andmeanspeedaccuracy(ml/h) | 1000±10% |
max volume flushandmeanspeed accuracy(ml) | 40±10% |
max feeding pump pressure | ≥0.1 MPa |
occlusion alarm level (pressure) | 50KPa±10KPa |
the longest occlusion triggering time of this device at minimum flow rate | 43min |
the shortest occlusion triggering time of this device atintermediate flow rate | 85s |
Medium speed running the pills from blocking the alarm threshold | 0.79ml |
recoverable Alarm | 1min50s—2min |
pause overtime alarm time | 1min50s—2min |
heating temperature | Infusion flow rate≥ 20ml / h, 15cm distance away from the heater outlet at 25-38 ℃ (EA312/313 models have this feature) |
use ofinfusionset | see recommended infusion sets list |
air-in-line sensitivity | detestability of single size of air bubble ≥50uL |
alarm | occlusion alarm, Over heatedalarm, battery depleted alarm |
alarms | air in line,Prime, Flush, low power, failurecodes (ERxx), pause overtime, powerstate, infusion mode, work state, heaterstate. |
noise | measured at the highest flow rate on wooden table, the volume is lower than 50 dB at a distance of 1 meter |
power | ac: 100V-240V 50 Hz /60Hz |
built inbattery | a fully charged internal DC11.1V battery can continuously run the pump for approximately 5 hours on intermediate rate |
power consumption | ≤40VA |
power output | ≤20VA |
The rated output voltage | 15V |
weight | 2.0kg |
heater | W-E01 ,cable length is 1.5m,Shield |
Essential performance | This product is suitable for patients with enteral nutrition infusion, the host cancontrol the infusion rate. With alarmfunction, with bubble detection piping and piping blockage detection. |
classification | type BF,class ΙΙ continuous volumetric feeding pump with internal battery; IPX1 (sealed drip-proof device) |
dimensions | 155mm(L)×170mm(W)×180mm(H) |
accessories | 1 user manual, 1 power cable, 1 heater (EA312/313 models only) |
operational | operating temperature: 5℃~+40℃ |
relative humility: 20%~90% | |
conditions | atmospheric pressure: 700hPa~1060hPa |
feeding pump should be operated in conditions free from strong shocks or vibrations, strong magnetic fields or corrosive gas environments. | |
battery operating condition | charging:0 ℃~+45 ℃ discharge:-20℃~+60℃ |
servicelife | 5 years |