Bioevopeak 120 Tests/Hour, Double Reagents Fully-auto Biochemistry Analyzer

Share on:


Price:$3,650.00 - $3,873.00
Price in USD:Details

Quantity:


Product Overview

Description


111


Fully-auto Biochemistry Analyzer, BA-A-120


 



Features


 



  • Constant 120 test / hour with double reagent

  • 80 reagent positions

  • Automatic liquid level sensing, anti-collision protection.

  • Integrated full closed optical system, multi wavelength simultaneous detection.

  • New amorphous optical plastics reaction cups, perfect transmittance.

  • Maintenance free semiconductor solid state refrigeration

  • Friendly surface and easy to operation


  • Multifunctional precision sampling probe



    • External & internal mirror polish, external & internal probe washing

    • Dedicated sampling probe equipped with sensitive liquid sensor, timely feedback of reagent & sample residuals

    • Collision protection, automatic probe depth adjusting




  • Excellent mixer design



    • Teflon coating mixer. No water dropping(reducing carry-over)

    • Excellent mixing effect with standard mixing prosedure




  • High precision ceramic syringe



    • Permanent ceramicpiston

    • Accurate dispensing as low as 0.1µL




  • Reagent tray & Sample tray;



    • 80 reagent positions,40 sample positions

    • Non-stop cooling system with peltier pad inside, 24 hours 2℃-14℃

    • Bar code reader(optional)




  • Reaction tray



    • 50 reaction cuvettes

    • Reaction volume as low as 150μL

    • Stable & accurate temperature (37±0.1℃)for reaction




  • Advanced operating system



    • User friendly software interface simple, easy-to-operation

    • Sample automatic dilution(decrease, increase, normal)

    • Reaction result automatic checking and warning, calibaration result automatic checking, warning alarm log

    • Linear & nonlinear calibration, Multi-points calibration up to 8 points

    • Quality control apply to westgard, L-T plot, cumulative







Specifications


 

























































































































































General information
Machine typeRandom access, open reagent & locked reagent (optional)
Test speedConstant 120 tests/hour (Mono reagent or double reagent)
Test principleColorimetric method, turbidimetry
Test method1 point end, 2 point end, Fixed-Time, Kinetics
Calibration typeLinear & nonlinear
Sample unit
Sample tray40 sample positions
Sample tubeMicro cup & test tube & blood collection tube can be used. (Φ 12 -13 )mm *(25-100 )mm
Sample volume20-300uL step by 0.1uL
Sample probeLiquid surface detection, timely tracking for liquid volume. Collision avoidance
Sample barcodeSample barcode reader (optional)
Reagent unit
Reagent tray80 reagent positions (40 for Rl & 40 for R2), Non-stop cooling system with peltier pad inside, 24 hours 2°C-14°C, open system, could accept any reagent
Reagent volume20-300uL, step by luL
Reagent needleLiquid surface detection, timely tracking for liquid volume. Collision avoidance
Regent barcodeRegent barcode reader (optional)
Reaction unit
Reaction tray50 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance.
Reaction volume150uL-330uL
Reaction time10 minutes
TemperaturePeltier pad incubation system, reaction temperature (37°C ±0.1 °C)
Mixing system
Mixerpaddle-type mixing, effective reducing of carry-over
Optical system
Light sourceHalogen lamp
Wavelength340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm;700nm(4 more options)
Absorbance range0-4.0 Abs
Resolution0.0001 Abs
Operating system
Operation systemWindows XP, Windows 7, Windows 8, Windows 10
Data processingAutomatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory, cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification, user operation privilege classification management.
Report printing6 formats optional, surpport custom pattern
interfaceTCP/IP network interface
Working conditions
Power supplyAC 110/220V ± 10%, 50/60 Hz, 650W
Water consumption6L/hour(maximum)
Dimension750mm*550mm *490mm(L*W*H)
Weight

70kg


 






Packaging & Shipping


sss 





About us


-(1)(1)-01 





Certifications


fff 





FAQ


Q: How about quality of the goods?


A:All the running products comply with CE standards.


Independent quality inspection team, conducting multiple inspections and random inspections from parts supply to product dispatch.


Well-known third-party testing service available, such as SGS, BV, and TUV, etc.


 


Q: What are your pre- and after-sales services?


A:Timely reply to pre-sales enquiries


Real-time update of manufacturing and shipment status


Contracted third-party maintenance company in over 60 countries


Assured availability of parts supply for 10 years at least


 


Q: Is OEM service available.


A: Most of the running products are available for OEM customization, including logo, package and function, if attainable, customization.


 


Q: How long is the lead time:


A: Generally, the lead time for most stocked products is about 7~15 days, vary by models and production schedule. For those non-stocked products, the manufacturing lead time is around 20 days.


eee 