Share on (1600447679647):



Principle of Operation | Non-dispersive infrared(NDIR) single beam optics, dual wavelength, no moving parts. |
Initialization Time | Capnogram displayed in less than 3 seconds,full specifications within 3 minutes(At 25°C). |
CO2 Measurement Range | 0 - 114 mmHg; 0 - 15 %; 0 - 15.2 kPa. |
Sampling Frequency | 100Hz |
Respiration Rate Range | 3 to 150 breaths per minute. |
Respiration Rate Accuracy | ±1 breath. |
Calibration | No routine user calibration required. An airway adapter zero is required when changing to a different style of airway adapter. |
Airway Adapters | Single-patient-use or reusable, ≤6cc dead space (Adult/Pediatric), ≤1cc dead space (Neonatal). |
Temperature | Operating: 0 to 40°C; Storage: -40 to 70°C; Transport: -40 to 70°C. |
Humidity | Operating: 10 to 90% RH, non-condensing; Storage: 10 to 90% RH, non-condensing. Transport: 10 to 90% RH, non-condensing. |
Atmospheric Pressure | Operating: 400-800 mmHg; Storage: 400-800 mmHg; Transport: 400-800 mmHg. |
Water Resistance | IPX4 - Splash-proof - Sensor only. |
Voltage Requirements | 5VDC ± 5 % |
Power Rating | <1.0W Typical; <1.8 W Warm-up. |
Sensor Weight | <85g. |
Sensor Size | Deep*Wide*High:48mm * 24.3mm * 35.5mm. |
Cable Length | 2.4m. |