Share on (1600465633061):

1 | Application field: | The electrode is an electrochemical sensor for measuring the concentration of nitrate ions in solution | ||
2 | Linear range: | 100~5×10-5M | ||
3 | pH range: | nitrate ion concentration pH 1×10-1M 2.5~10.0 1×10-3M 3.5~8.5 | ||
4 | Response time: | 5×10-5M <2 minutes 100~10-4M <1 minute | ||
5 | Electrode internal resistance: | <1Ω (25℃) | ||
6 | Main interfering ions: | Anion | selection factor | |
Chloride | 4.9×10-2 | |||
Citrate | 7.3×10-5 | |||
Sulfate ion | 4.1×10-5 | |||
Tartrate | 1.4×10-4 | |||
Dihydrogen phosphate ion | 1.1×10-4 | |||
Perchlorate ion | Obvious interference | |||
Hydrogen phosphate ion | 1.4×10-4 | |||
Bromide ion | interference | |||
Hydrogen acetate ion | 2.7×10-4 | |||
Iodide ion | interference | |||
EDTA disodium salt | 5.7×10-5 | |||
Ammonium ion,/potassium ion/ferric ion | great influence | |||
7 | Advantage: | The electrode can be used independently without the need for a reference electrode | ||
8 | Dimensions: | 160mmX16mm (diameter) | ||
9 | Weight: | 55g | ||








New products from manufacturers at wholesale prices