Share on (1600520246258):




Parameter | OP-PCR1600C | Other Similar Products | Comparison Description | ||
Lowest Detection Range | 10 copies/ml | 200-500 copies/ml | Opulen Real-time fluorescent PCR analyzer has higher sensitivity and accurate detection to weak positive samples. | ||
Temperature Uniformity | ≤±0.1°C (@58°C) | ≤±0.2~0.5°C | Opulen PCR analyzer has more precise temperature control and high uniformity. It is conducive to the effective amplification of samples and ensures the detection accuracy of all samples. | ||
Temperature Accuracy | ≤±0.1°C (@58°C) | ≤±0.2~0.5°C | Opulen PCR analyzer has more precise temperature control and high uniformity. It is conducive to the effective amplification of samples and ensures the detection accuracy of all samples. | ||
Average Heating Rate | ≥8.0℃/sec | ≥3.5~6.0℃/sec | Opulen PCR analyzer, the heating and cooling speed is faster, which can speed up the effective amplification of samples and shorten the detection time. | ||
Average Cooling Rate | ≥5.0℃/sec | ≥2.0~4.0℃/sec | Opulen PCR analyzer, the heating and cooling speed is faster, which can speed up the effective amplification of samples and shorten the detection time. | ||
Display Screen | Colorful 10.1 inch touch type screen | 7.0-10 inch | Opulen PCR analyzer has larger and more intuitive screen with more portable operation. | ||
No. | Basic Performance | Parameter | ||||
1 | Sample capacity | Support 16 wells with 0.2mL | ||||
2 | Reaction system | 30-100μL | ||||
3 | F1:FAM, SYBR GreenI, etc F2: HEX, VIC, JOE, etc F3: ROX, TEXAS RED, etc F4: Cy5, etc | |||||
4 | Temperature range | 30-100℃ | ||||
5 | Temperature control | Heating rate | ≥3.5℃/s | |||
Cooling rate | ≥3.5℃/s | |||||
Temperature fluctuation | ≤0.2℃ | |||||
Temperature indicating value error | Difference absolute value ≤0.5℃ | |||||
Temperature uniformity | Temperature difference : ±0.1℃ | |||||
Temperature duration accuracy | Relative deviation: ±5% | |||||
6 | Fluorescence intense repeatability | Relative standard deviation≤2% | ||||
Fluorescence intense uniformity | Relative standard deviation≤5% | |||||
Fluorescence intense linearity | Linear regression coefficient r ≥0.990 | |||||
7 | Confidence level | ≥99.9% | ||||
Device linearity | Linear regression coefficient r ≥0.99 | |||||
Quantitative repeatability | Relative standard deviation≤3% | |||||
Quantitative indicating value error | Max. Value ≤ 15% | |||||
8 | Display screen | 10.1 inch LCD Touch screen | ||||
9 | Power | 100-240VAC 50/60Hz Max1200VA | ||||
10 | Interface | USB port, serial port | ||||
11 | Dimension | 530mm× 530mm×380mm (L*W*H) | ||||
12 | Weight | 10.8Kg | ||||