Basic Characteristics
|
Reaction temperature 37.0℃
|
Control accuracy of temperature ±1.0℃
|
Test items At most 8 items can be simultaneously tested.
|
Test methods Clotting assay, chromogenic substrate assay, and immunoturbidimetric assay
|
Test speed PT single-item test: 100 samples/ hour
|
STAT priority function
|
Reagent System
|
Reagent chamber, reagent position:
|
With refrigerated reagent chamber, 20 refrigerated reagent positions
|
Normal-temperature reagent rack, 3 normal-temperature reagent positions
|
Reagent volume 20μL~175μL
|
Specifications of reagent bottle 5mL, 15mL, and 50mL
|
Storage temperature of reagent Refrigerated reagent ≤16.0℃
|
Automatically calculate the remaining volume and test numbers
|
Automatic barcode scaning to recognize the reagent information
|
Reagent liquid level sensor Integrated with probe
|
Data System
|
Port: Network port
|
Computer configuration: Optional computer and printer
|
Connection with LIS/HIS system: It is supported
|
Analysis System
|
Incubation and test positions: Totally 16, 8 incubation positions and 8 test positions
|
Volume of cuvette: 600 μL
|
Optical system: Light-emitting diode, photodiode
|
Abnormal sample can be diluted and re-run automatically
|
On-board adding reagents,samples and cuvettes,work effectively
|
Reagent probe heating function.
|
Non-contact mixing method
|
QC: Real-time QC, daily QC and inter-day QC with multi-concentration method
|
Complete Machine System
|
Weight: 75kg
|
Power consumption: 450VA
|
Outline dimensions: 660mm×475mm×554mm(L×W×H)
|