Share on (1600570200643):

Specification | ||
Parameters | ECG, HP, RESP, SPO2, NIBP,TEMP | |
Product Size | 210*85*180mm N.W.:1.35kg | |
Packing Size | 263*255*190mm N.W.:3.4kg | |
Apply For | Adult/Pediatric/Neonatal | |
Application | Medicine/Surgery/Operating Room/ICU/CCU | |
Display | 8.4 inch LCD Screen | |
Operating Environment | ||
Temperature: | 0 ~ 40 ℃ | |
Humidity: | ≦85% | |
Pressure: | 86.0 kPa ~ 106.0kPa | |
Transport and Storage environment : | ||
Temperature: | -20 ~ 55 ℃ | |
Humidity: | ≦93% | |
Pressure: | 86.0 kPa~ 106.0kPa | |
Power Requirements | ||
AC: | 100 ~ 240V, 50Hz/60Hz | |
DC: | Built-in rechargeable battery | |
Battery: | 11.1V 24wh Li-ion battery; 2h operating after full charge; 5min operating after low battery alarm; | |
Date Storage | ||
Trend diagram/table: | 720h | |
NIBP review: | 10000 events | |
Wave review: | 12h | |
Alarm review: | 200 alarm events | |
Analysis: | Support drug concentration titration analysis | |
ECG | ||
5 Leads: | RA, LA, LL, RL, V | |
Lead mode: | I, Ⅱ, Ⅲ, AVR, AVL, AVF, V | |
Gain: | × 1.25mm/mV,2.5mm/mV, 5.0mm/mV,10mm/mV, 20mm/mV (error ≤ 5%) | |
Sweep speed: | 6.25mm/s,12.5mm/s,25mm/s, 50mm/s (error ≤ ± 10%) | |
Bandwidth: | Surgery 1 ~ 25 Hz, Monitor 0.5 ~ 40 Hz, Diagnostic 0.05 ~ 130 Hz | |
Sensitivity: | >200 u V P-P | |
CMRR: | ≥ 112dB | |
Differential Input Impedance: | >5M | |
ohm Electrode Offset Potential: | ±300mV | |
Polarization Voltage: | ≥ ± 500mV | |
Input Impedance: | ≥ 5 M | |
Calibration Voltage: | 1mV ± 5% | |
ST Segment | ||
ST Segment Range: | -2.0mV ~ +2.0mV | |
Accuracy: | 0.02mV | |
Optional accessories | 1.Trolley bracket | |
2.Hanging bracket | ||
3.Monitor recorder | ||
4.Vehicle-mounted power supply | ||
If you need more product parameter,please send an inquiry to us~ | ||