Share on (1600601137709):

Product name | Heal Force Electrosurgical Unit 350W | ||||||
Rate frequency | 512 kHz (110/240vAC, 60/50Hz) | ||||||
Isolated output | YES | ||||||
Quantitative display of contact area | YES, for "Split" type of neutral electrode | ||||||
Contact area indicator lamp | 10 lamps (one for 10% contact), alarm beep and flicker if contact < 20% | ||||||
Split/solid neutral electrode definition | YES, "Split" or "Solid" indicator lamp lights up | ||||||
Adjustable alarm volume | YES | ||||||
Operational duty cycle | 10 seconds ON, 30 seconds OFF | ||||||
Cooling method | Convection | ||||||
Accessory & supply connection | Finger-switch pencil | 1 | |||||
Finger-switch pencil | 2 | ||||||
Patient return electrode | 1 | ||||||
Foot-switch | 1 | ||||||
AC power supply | 1 | ||||||
Equipotential bonding | 1 | ||||||
Surgical smoke evacuator | 1 (but surgical smoke evacuator is optional) | ||||||
Regulation | Type according to EN 60601-1 | CF | |||||
Protection class to EN 60601-1 | class I | ||||||
Protection class to EN 60601-1 | IIb | ||||||
Packaging | Dimension (WxDxH) | 330 x 450 x 150 mm | |||||
Weight | 7.0 KG | ||||||
Monopolar | ||||||||
Cutting mode | Max. power output (watts) | Load (ohms) | Open circuit Vp-p | Crest factor | ||||
Pure Cut | 350w (grey) / 150w (white) | 500 Ω | 1,800 / 1,200v | 1.5 | ||||
Blend1 | 300w (grey) / 120w (white) | 500 Ω | 2,700 / 2,000v | 1.9 | ||||
Blend2 | 200w (grey) / 80w (white) | 500 Ω | 2,400 / 1,700v | 2.3 | ||||
Blend3 | 150w (grey) / 60w (white) | 500 Ω | 2,900 / 3,000v | 3.1 | ||||
Coagulation mode | ||||||||
Point | 100w (grey) / 70w (white) | 500 Ω | 4,600 / 3,900v | 4.5 | ||||
Soft | 100w (grey) / 70w (white) | 500 Ω | 3,600 / 3,300v | |||||
Bipolar | ||||||||
Cutting mode | - | - | - | - | ||||
Standard | 50w (grey) / 50w (white) | 100 Ω | 600V | 1.5 | ||||











New products from manufacturers at wholesale prices