Item name | steam sterilization packaging machine | ||||||
Water capacity | External pure water | ||||||
Saturated steam temperature | 102°C | ||||||
Rated voltage | 110-220V | ||||||
Rated frequency | 50HZ | ||||||
Rated input power | 2.5KW, 3KW | ||||||
Size | 480*400*320mm | ||||||
Weight | 18.25kg | ||||||
Common malfunctions | Failure phenomenon | Approach | ||
1 | No pumping after turning on the power switch (the electromagnetic pump does not sound) | The electromagnetic pump is faulty or the power is not turned on, replace the electromagnetic pump | ||
2 | The equipment is working, and the amount of steam is large, sometimes small or long The temperature won't rise | 1. It may be that the thermostat is disconnected, wait for 10 minutes to reduce the temperature and restart, if it still does not It says that the thermostat is damaged and needs to be replaced. 2. Check whether the steam nozzle and heater are blocked by foreign objects, and they still cannot be replaced after the blockage is cleaned. Heater. | ||
3 | No water is pumped, but the electromagnetic pump makes a noise | 1. The steam nozzle is blocked, clean the scale until the steam nozzle is unblocked. 2. It may be caused by water shortage, check whether the water supply is normal. | ||
The above failure types are for reference only. If any failure occurs, stop using it immediately and notify the professional maintenance personnel to repair it as soon as possible. | ||||