Share on (1600698108944):


Specification: | ||||||||
1 | Basic Requirement | Laminar Air Flow should be Ceiling Mounted with 99.999% HEPA Filter (0.3 micron) efficiency according to cGMP to avoid particulate & microbial contamination during filling process. It should comply DOP test and air velocity criteria as per WHO/MHRA guideline. | ||||||
2 | Dimensions | Customized | ||||||
3 | Type | Floor Mounted(With Stand fram) | ||||||
4 | MOC | SS304 | ||||||
5 | Noise Level | 60 to 65 db from 1 meter distance | ||||||
6 | Blower Assembly | Statically & Dynamically balanced with suitable rating & size | ||||||
7 | Air Velocity | 0.45 +/- 20% m/sec under HEPA Filter | ||||||
8 | DOP Port | DOP/PAO port should be provided | ||||||
9 | Class | Class-100 | ||||||
10 | PVC Curtain | Three side pvc curtain,one side open, curtain length 1850 mm down from the LAF | ||||||
11 | Air Flow Pattern | Vertical down flow | ||||||
12 | Documentation | Operation & Maintenance manual should be provided. Drawing of the equipment showing front elevation, side elevation & plan view should be provided. IQ and OQ dully prepared and supplied. | ||||||
13 | Warranty | one year waranty from the date of commissioning. | ||||||


















New products from manufacturers at wholesale prices