Power requirements | Voltage | AC380V/400V±10% |
Frequency | 50Hz | |
Resistance | ≤0.1Ω | |
Capacity | ≥63kVA | |
Mechanical parameter | Column vertical movement distance | 1160mm (Error: ±5%) |
Column lateral movement distance | 1730mm (Error: ±5%) | |
Chest film frame lifting distance | ≥1200mm | |
Tube assembly rotation angle | -90°~+90°, manual rotation | |
Column rotation angle | -180°~+180°, manual rotation | |
Patient table X-axis movement distance | 900mm (Error: ±5%) | |
Patient table Y-axis movement distance | 260mm (Error: ±5%) | |
Patient table dimensions (L×W×H) | 2110mm×822mm×656mm (Error: ±5%) | |
Max. Patient Weight (kg) | 150kg | |
Effective filtration of photographic table | 0.2 mmAl | |
High voltage generator | Input power voltage | AC 380V/400V±10% |
Frequency | 50Hz | |
Output nominal electric power | 50kW (500mA,100kV,0.1s) | |
Tube voltage/kV | 40kV~150kV, digital button adjustment, minimum adjustment step length is 1kV | |
Tube current/mA | 10mA~630mA, meet the value of R'10 number system in ISO 497, adjusted in a multilevel manner by digital button | |
Loading time | 1ms~6300ms, meet the value of R'10 number system in ISO 497, adjusted in a multilevel manner by digital button; AEC nominal shortest irradiation time: 1.6ms | |
Current time product/mAs | 0.1mAs~630mAs, meet the value of R'10 number system in ISO 497, adjusted in a multilevel manner by digital button |