Main technical parameters and working environment requirements:
| Product Name | Disposable IBP Transducer for Utah / Utah Blood Pressure Transducer |
| Barometric pressure | 70~106Kpa |
| Relative humidity | 10~90% (non-condensing) |
| Operating pressure range | -50 ~ + 300 mmHg |
| Sensitivity | 5.0μV/V/mmHg ±3% |
| Nonlinearity and hysteresis | full-scale reading±1.5% |
| Input impedance | 1200Ω ~3200Ω |
| Output impedance | 300±5% |
| Zero pressure offset | -20mmHg~+20mmHg |
| Thermal offset shift | ≤±0.3mmHg/°C |
| Offset drift | After warm-up for 20 seconds, drift within 2 mmHg within 8 hours |
| Thermal span shift | ≤±0.1%/°C |
| Frequency response | Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz |
| Defibrillator withstand | Defibrillation depends on the final connection of the equipment |
| Leakage current | Leakage current is dependent on the final connection of the equipment |
| Overpressure load | -400~+4000mmHg |
| Shock resistance | Shock resistance With stand falling from one meter for three times |
| Light sensitivity | When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage |
| Contact with the human body | ≤168h |
| Storage and transport environment requirements | - 20 ° C~ + 60 ° C, <90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold. |
Intra-system velocity Flow 2-5mL/h Intra-system velocity refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions | |
| Product shelf-life | 3 years |