Channel | Volume Range | Increment | Test Volume | Error limits in accordance to ISO8655-2 | |||||||||
accuracy error | precision error | ||||||||||||
μL | μL | μL | μL | % | s.d.*μL | CV*% | |||||||
1 | 0.5-10 | 0.01 | 10 | ±0.10 | ±1.00 | 0.05 | 0.5 | ||||||
1 | ±0.035 | ±3.50 | 0.03 | 3 | |||||||||
1 | 5-50 | 0.1 | 50 | ±0.40 | ±0.80 | 0.15 | 0.3 | ||||||
5 | ±0.15 | ±3.00 | 0.125 | 2.5 | |||||||||
1 | 30-300 | 1 | 300 | ±1.80 | ±0.60 | 0.6 | 0.2 | ||||||
30 | ±0.90 | ±3.00 | 0.21 | 0.7 | |||||||||
1 | 100-1000 | 5 | 1000 | ±6.00 | ±0.60 | 2 | 0.2 | ||||||
100 | ±3.00 | ±3.00 | 0.6 | 0.6 | |||||||||
* s.d. = Standard Deviation | Specifications are used as guidelines and the user calibration should refer |