UV-visible spectrophotometer | |||||||
Model | L7 | L8 | L9 | 760CRT | |||
Metering method | Dual beam | ||||||
Monochromator | Czerny - Turner | ||||||
Focal length | 160mm | 200mm | |||||
Grating | 1200 lines/mm | 1600 lines/mm | Holographic sparkle, 1600 lines/mm | ||||
Beam splitting system | - | chopper | |||||
Detector | silicon photovoltaic cells | Imported receiver | Imported photomultiplier tube R928 | ||||
Spectral bandwidth | 0.08~5nm, interval 0.01nm, continuously adjustable | ||||||
Wavelength setting | Touch screen input | computer input | |||||
Wavelength range | 190-1100nm | 190-900nm | |||||
Wavelength accuracy | ±0.3nm | ±0.3nm(actual measurement≤0.2nm) | ±0.3nm (built-in automatic wavelength correction device) | ||||
Wavelength repeatability | <0.2nm | <0.1nm | <0.2nm | ||||
Resolution | - | Better than 0.15nm | |||||
Wavelength scan speed | fast, medium, slow | ||||||
Light source switching wavelength | 340nm | ||||||
Stray light | <0.03%(T) | <0.02%(T) | <0.08%(T) | ||||
Photometric range | 0.0~200.0% T -0.301~4.000A 0.000~9999C | ||||||
Photometric accuracy | ±0.3%T | ±0.3%T ±0.002Abs(0-0.5A) ±0.004Abs(0.5-1A) | ±0.3%T (measured with NBS930D) | ||||
Photometric Repeatability | <0.15%T | <0.15%T 0.001Abs(0-0.5A) 0.002Abs(0.5-1A) | <0.2%(T) | ||||
Baseline straightness | ±0.0015A(200~1090nm) | ±0.0008A | ±0.002A | ||||
Noise | 100% (T)≤0.15%(T) 0% (T) ≤ 0.1%(T) | 0.1%T | ±0.3(T)(100% position)±0.1%(T)(0% position) | ||||
Baseline drift | ±0.0009Abs/0.5h (2h after power-on, at 500nm) | ±0.0009Abs/0.5h (2h after power-on, at 250nm and 500nm) | <±0.004 A/30min | ||||

UV-visible spectrophotometer | ||||||||
Model | L3/L3S visible | L5/L5S | L6/L6S | 757 | ||||
Metering method | proportional Dual beam | |||||||
Monochromator | Czerny - Turner | |||||||
Focal length | 160mm | |||||||
Grating | 1200 lines/mm | |||||||
Detector | Imported receiver | |||||||
Spectral bandwidth | 1.8nm | |||||||
Wavelength setting | Touch screen input | |||||||
Wavelength range | 325-1100nm | 190 -1100nm | ||||||
Wavelength accuracy | ±0.5nm | ±0.3nm | ±0.5nm | |||||
Wavelength repeatability | ≤0.2nm | ≤0.1nm | ≤0.2nm | |||||
Wavelength scan speed(S type) | fast, medium, slow | - | ||||||
Light source switching wavelength | 340nm | |||||||
Stray light | ≤0.05% (T) | ≤0.05% (T) | 0.03% (T) | ≤0.05% (T) | ||||
Photometric range | 0.0~200.0% T -0.301~4.000A 0.000~9999C | |||||||
Photometric accuracy | ±0.5%T ±0.008Abs(0.5-1A) ±0.004Abs(0.5-1A) | ±0.3%T ±0.004Abs(0-0.5A) ±0.002Abs(0-0.5A) | ||||||
Photometric Repeatability | ≤0.2%T 0.004Abs(0.5 -1A) 0.002Abs(0.5 -1A) | ≤0.15%T 0.002Abs(0-0.5A) 0.001Abs(0-0.5A) | ||||||
Baseline straightness(S model) | ±0.002A | ≤±0.002A (200nm~1090nm) | ≤±0.0015A (200nm~1090nm) | - | ||||
Noise | 100% (T)≤0.2%(T) 0% (T) ≤ 0.1%(T) | 0.15%T | 100% (T)≤0.15%(T) 0% (T) ≤ 0.1%(T) | |||||
Baseline drift | ±0.0009Abs/0.5h (2h after power-on, at 500nm) | ±0.0009Abs/0.5h (2h after power-on, at 250nm and 500nm) | ||||||