Specification | ||||||||
Measurement function | 1.3 Measurement function 1.3.1 Output urine flow rate curve: The urine flow rate range is 0-30ml/s, the time is 20 seconds, 40 seconds, and 60 seconds, and the gears are automatically switched. 1.3.2 The output parameters are: total urine volume, urine flow time, average urine flow rate, large urine flow rate, large urine flow time, time to reach 90% of urine volume, time to reach 10% of urine volume, total urination time, etc. 8 items, data and graphics can be output through the built-in printer 1.3.3 Automatic calibration "0" function: This machine automatically calibrates each parameter to "0" before entering measurement work. The influence of residual urine or urine cup weight in the early stage was eliminated. *1. 3. 4 Urine collection system scope: The urine collection bag has a large diameter of 30 cm and a height of 50 cm; the urine bag holder (ring) has a diameter of 16 cm, is made of stainless steel, and is designed to be integrated with the host box. *1. 3. 5 Comes with disposable smart consumables to avoid cross-infection. *1. 3. 6 The unique pulley design on the sole of the foot makes it easy to move when measuring. 1.4 Measurement performance 1.4.1 Measurement accuracy: 10ml-50ml, when 50ml is not included, it should be ≤ soil 1ml; within 50ml tov700ml, when 50ml is included, it should be ≤ soil 2%. Waiting time for urination: System. |