Performance | ||||
ECG | Lead mode | 3-lead or 5-lead | ||
Lead selection | I, II, III, aVR, aVL, aVF, V | |||
Waveform | 5-lead, 2-channel3-lead, 1-channel | |||
Gain | 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV | |||
Scan speed | 12.5mm/s, 25 mm/s, 50 mm/s | |||
HR | Measurement and alarm range | 15~350bpm | ||
Accuracy | ±1% or ±1bpm, whichever is greater | |||
Alarm accuracy | ± 2 bpm | |||
Resolution | 1 bpm | |||
Measurement and alarm range | -2.0mV~+2.0mV | |||
Accuracy | 0.8mV~+0.8mV ±0.04mv or ±10%, whichever is greater | |||
Other range | unspecified | |||
Arrhythmia analysis | ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC | |||
Pacemaker | yes | |||
RESPIRATION | Method | R-F(RA-LL) Impedance | ||
Measurement and alarm range | 0~150rpm | |||
Resolution | 1 rpm | |||
±2 rpm | ||||
Alarm accuracy | ±3 rpm | |||
Apnea alarm | 10~40s | |||
Scan speed | 6.25 mm/s, 12.5 mm/s, 25 mm/s | |||
NIBP | Method | Oscillometry | ||
Mode | Manual/AUTO/STAT | |||
Measurement interval in AUTO mode | 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes | |||
Measurement period in STAT mode | 5 minutes | |||
Measurement and alarm range | 10 ~ 270 mmHg | |||
Resolution | 1 mmHg | |||
Cuff pressure accuracy | ±3 mmHg | |||
Maximal mean deviation | ±5 mmHg | |||
Maximal standard deviation | 8mmHg | |||
Adult mode | 297±3 mmHg | |||
Pediatric mode | 240±3 mmHg | |||
Neonatal mode | 147±3 mmHg | |||
SpO2 | Measurement and alarm range | 0 ~ 100% | ||
Resolution | 1% | |||
Measurement accuracy | 70%~100%: ±2%0%~69%: unspecified | |||
PR | Measurement and alarm range | 30 ~ 250 bpm | ||
Measurement accuracy | ±2 bpm or ±2%, whichever is greater | |||
TEMP | Channel | dual-channel | ||
Measurement and alarm range | 0 ~ 50℃ | |||
Resolution | 0.1 ℃ | |||
Accuracy | ±0.1 ℃ | |||

Accessories | veterinary tongue-clip SpO2 sensor (1M) | ||||
SpO2 sensor extension cable (2M) | |||||
Neonatal capsule single tube repeatable cuff (6-11cm) | |||||
NIBP extension tube | |||||
ECG lead cable | |||||
ECG veterinary clip | |||||
Transvaginal TEMP probe | |||||
Power cord | |||||
User Manual | |||||
Physical characteristic | Dimension | 235 mm(L) × 141 mm(W) × 222 mm(H) | |||
Weight | 2.5 Kg | ||||