Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Control Mode: Pre-warm Mode, Manual Mode and Baby Mode are controlled by micro-computer Baby Mode temperature control range: 34.5-37.5℃
Skin temperature display range: 5℃-65℃
Accuracy of skin temperature sensor: ±0.2℃
Temperature uniformity of mattress: ≤2℃
Dominant wavelength of blue light: 400-550nm
Timing range of blue light: 0~9999hrs 59min
APGAR timer: audible and visible alarm when the device runs to 50″~1′,4′50″~5′,9′50″~10′
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
Average value of general bilirubin radiation on the mattress effective range: ≥250uW/cm2(the upper phototherapy equipment)
≥600uW/cm2 (the nether phototherapy equipment)
Bilirubin uniformity of radiating on the mattress effective range: >0.4
Environmental Requirement
Operating range: 18℃~30℃
Ambient air velocity rate: <0.3m/s
Product Packaging
Package: The whole unit is packed into one carton
Dimension: 940mm x 570mm x 1080mm
Gross Weight: 57kg
Product advantage:
Three control Modes: Pre-warm Mode, Manual Mode and Baby Mode control by micro-computer;
Set Temperature and skin temperature can be displayed separately;
Main body is made of stainless steel;
Double sides phototherapy radiant box;
Acrylic glass bassinet;
APGAR timer;
Multiple failure alarm indications;
RS232 connector for data output;
Disposable skin temperature sensor and SpO2 are optional.

Three control Modes: Pre-warm Mode, Manual Mode and Baby Mode control by micro-computer;
Set Temperature and Baby Temperature can be displayed separately;
Horizontal angle of warmer module and inclination of the bassinet can be adjusted;
The panels surrounded bassinet can be turned outward or disassembled;
X-ray cassette under the bassinet;
APGAR timer;
Multiple failure alarm indications;
RS232 connector for data output.