kinematic viscosity testing apparatus/ oil kinematic viscosimeter/ capillary viscometer
The instrument is designed and made as per the T0619 "Asphalt Kinematic Viscosity Test (Capillary Viscometer Methods". It is suitable to determine kinematic viscosity of asphalt at a certain temperature.
Power supply: AC 220V± 10%, 50 Hz;
| Heating power: | Two grades; 1000 W for auxiliary heating and 600 W for temperature controlling |
| Bath temperature | Ambient temperature~180.0ºC |
| Temperature controlling accuracy | ± 0.1ºC |
| Mercury thermometer | 1) Rod type (2) Scale division: 0.1 ºC; |
| Bath cubage | Not less than 23 L |
| Sample quantity | 3 piece at the same time |
| Stirring motor | 6 W, 1200 RPM |
| Ambient temperature: | -10ºC~+35ºC |
| Relative humidity: | <85% |
| Temperature sensor: | Industrial platinum resistance, PT100 |
Total Power consumption: Not more than 1800 W
Capillary viscometer: 7 pieces of Cannon-Fenske reserveflow capillary viscometer tubes ina group; Their models are 200, 300, 350, 400, 450, 500, 600 respectively (Corresponding diameters are 1.02, 1.26, 1.48, 1.88, 2.20, 3.10 and 4.00 mm). The users should prepare Ubbelohde type viscometer tubes by
themselves.