GC 112 GC122 Gas chromatograph with FID or TCD detector

Share on (1992802051):


Price:

Quantity:


Product Overview

Description



Product Description


 Column Compartment/oven with superior thermal performance, multistage (10 ramps) programmed temperature control function. (supported by “control system”)
Advanced built-in data acquisition system , supporting real time instrument status monitoring, detection signal acquisition and PC control
Column oven accommodates up to 3 chromatographic columns, and supports quick heat-up and rapid cool-down with automated back-door opening. 450 to 50 in 6 min at 25 ambient)
Flexible sample introduction system: 3 sample injectors could be installed and operated simultaneously with independent temperature control.
High sensibility and stability detector.
2 independent and analog signals output.
Clarity software, compatible with GLP/FDA-21 CFR Part11 requirements and regulations. (electronic records and signatures)
Special analyzers, like RGA, NGA, PINDA, etc. according to ASTM standards


Superior performance with advanced, microcomputer-based temperature control system
High temperature accuracy (optimum ± 0.02), high reliability, and anti-interference
Self-diagnosis/ self-protection function (overheat protection, power-off protection, etc.)
Intuitive display of timing program, detector status, measurement range, current setting, etc


Parameters


 


 


























































































Model#



GC122



GC112



Description



EPC


 

Sample Introduction System



sample injector and evaporation chamber



Column Oven



Temperature range



Ambient temperature +7℃ ~ 400℃ (in 1℃ increment)



Temperature accuracy



± 0.02℃



Cooling time



400℃ to 50℃ in 6 min at 25℃ ambient



Programmed temperature setting



0.1℃~40℃/min (in 1℃ increment)



FID



Detection limit



≤3×10-12g/s (C16)



Drift



≤6×10-13A/h



TCD



Sensitivity



≥5000mV.mL/mg



Baseline noise



≤20µV



FPD



Detection limit



≤8×10-13g/s(P) ; ≤8×10-11g/s(S)



Drift



≤2×10-11A/30min



Baseline noise



≤5×10-12A



ECD



Detection limit



≤2×10-13g/s (γ-666)



Drift



≤80µV/30min



Baseline noise



≤20µV



Linear dynamic range



≥103



NPD



Detection limit



≤5×10-13g/s(P) ; ≤5×10-12g/s(N)



· 


 





Company Information


 JKI





FAQ


 


 


Q: How to buy your products ?


A: We have some products in stock , you can take away the products after you arrange the payment ; If we don’t have the products in stock you want, we will start  production once getting the payment.


 


Q: I need the equipments specification not list in your website, do you offer you the customized service.


A:Yes , we can offer the Customized service , based on clients requirements to manufacture the machine.


 


Q: What’s the warranty for the products ?


A: The free warranty is one year from the date of commissioning qualified.


 


Q: Can we visit your factory ?


A: Of course, welcome to visit our factory if you come to China. We can pick up at the airport.


 


Q: How long is the validity of quotation ?


A: Generally, our price is valid within one month from the date of quotation. The price will be adjusted appropriately as the price fluctuation of raw material in the market.


 


Q: What’s the production date after we confirm the order ?


A: This depends on the quantity. Normally, for the mass production, we need about one week to finish the production.




0.2111 s.