GUIDE WIRE HI TORQUE BALANCE MIDDLEWEIGHT ELITE

Share on (50035187107):


Price:

Quantity:


Product Overview

Description




IMPORTANT SAFETY INFORMATION
 




















HI-TORQUE™ Guide Wires



 



 

 









INTENDED USE


All HI-TORQUE™ Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).


INDICATIONS
Refer to the device label for any additional product-specific indications that may apply.


CONTRAINDICATIONS
HI-TORQUE™ Guide Wires Wires are not intended for use in the cerebral vasculature. Refer to the device label for any additional product-specific contraindications that may apply.


WARNINGS
This device is designed and intended for ONE-TIME USE ONLY. Do not resterilize and / or reuse.



Carefully observe the instructions under “Do Not” and “Do” below. Failure to do so may result in vessel trauma, guide wire damage, guide wire tip separation, or stent damage. If resistance is observed at any time, determine the cause under fluoroscopy and take remedial action as needed. Use the most suitable guide wire for the lesion being treated.


Do Not:



  • Push, auger, withdraw, or torque a guide wire that meets resistance.

  • Torque a guide wire if the tip becomes entrapped within the vasculature.

  • Allow the guide wire tip to remain in a prolapsed condition.



Do:



  • Advance or withdraw the guide wire slowly.

  • Use the radiopaque marker of the interventional device to confirm position.

  • Examine the tip movement under fluoroscopy before manipulating, moving, or torquing the guide wire.

  • Observe the wire under fluoroscopy for tip buckling, which is a sign of resistance.

  • Maintain continuous flush while removing and reinserting the guide wire to prevent air from entering the catheter system. Perform exchanges slowly to prevent air entry and / or trauma.

  • When reintroducing the guide wire, confirm that the interventional device tip is free within the vessel lumen and that the tip is parallel to the vessel wall.

  • Use extreme caution when moving a guide wire through a non-endothelialized stent, or through stent struts, into a bifurcated vessel. Use of this technique involves additional patient risks, including the risk that the wire may become caught on the stent strut.

  • Consider that if a secondary wire is placed in a bifurcation branch, this wire may need to be retracted prior to stent deployment because there is additional risk that the secondary wire may become entrapped between the vessel wall and the stent.




PRECAUTIONS

Guide wires are delicate instruments and should be handled carefully. Prior to use and when possible during the procedure, inspect the guide wire carefully for bends, kinks, or other damage. Do not use damaged guide wires. Using a damaged guide wire may result in vessel damage and / or inaccurate torque response.


Confirm the compatibility of the guide wire diameter with the interventional device before actual use.


Free movement of the guide wire within the interventional device is an important feature of a steerable guide wire system, because it gives the user valuable tactile information. Test the system for any resistance prior to use. Adjust or replace the hemostatic valve with an adjustable valve if it is found to inhibit guide wire movement.


Never attach the torque device to the modified portion of the proximal end of the extendable guide wire; otherwise, guide wire damage may occur, preventing the ability to attach the DOC™ Guide Wire Extension.


HI-TORQUE™ Guide Wires with Hydrophilic Coating: Avoid abrasion of the hydrophilic coating.


Do not withdraw or manipulate the hydrophilic-coated wire through a metal cannula or sharp-edged object.


ADVERSE EVENTS
Potential Adverse Events associated with use of this device may include the following but not limited to perforation, dissection, occlusion, myocardial infarction, embolism and infection.


 






















































































ProductPart NumberTip RadiopacityTip ShapeDistal CoatingsWire LengthMarker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™10118803 cmStraightHydrophilic190 cmMarker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™1011880J3 cmJHydrophilic190 cmMarker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™10118813 cmStraightHydrophilic300 cmMarker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™1011881J3 cmJHydrophilic300 cmMarker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™10118823 cmStraightHydrophilic190 cmNo Marker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™1011882J3 cmJHydrophilic190 cmNo Marker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™10118833 cmStraightHydrophilic300 cmNo Marker
HI-TORQUE BALANCE MIDDLEWEIGHT ELITE™1011883J3 cmJHydrophilic300 cmNo Marker





 







HI-TORQUE™ Guide Wires for PTCA, PTA, and Stents



 



rx-only-logo

 









INDICATIONS FOR USE: This HI-TORQUE™ Guide Wire is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). This guide wire may also be used with compatible stent devices during therapeutic procedures.


CONTRAINDICATIONS: Not intended for use in the cerebral vasculature or with atherectomy devices.



WARNINGS:


This device is designed and intended for ONE-TIME USE ONLY. Do not resterilize and / or reuse.



Carefully observe the instructions under “Do Not” and “Do” below. Failure to do so may result in vessel trauma, guide wire damage, guide wire tip separation, or stent damage. If resistance is observed at any time, determine the cause under fluoroscopy and take remedial action as needed. Use the most suitable guide wire for the lesion being treated.


Do Not:



  • Push, auger, withdraw, or torque a guide wire that meets resistance.

  • Torque a guide wire if the tip becomes entrapped within the vasculature.

  • Allow the guide wire tip to remain in a prolapsed condition.


Do:



  • Advance or withdraw the guide wire slowly.

  • Use the radiopaque marker of the interventional device to confirm position.

  • Examine the tip movement under fluoroscopy before manipulating, moving, or torquing the guide wire.

  • Observe the wire under fluoroscopy for tip buckling, which is a sign of resistance.

  • Maintain continuous flush while removing and reinserting the guide wire to prevent air from entering the catheter system. Perform exchanges slowly to prevent air entry and / or trauma.

  • When reintroducing the guide wire, confirm that the interventional device tip is free within the vessel lumen and that the tip is parallel to the vessel wall.

  • Use extreme caution when moving a guide wire through a non-endothelialized stent, or through stent struts, into a bifurcated vessel. Use of this technique involves additional patient risks, including the risk that the wire may become caught on the stent strut.

  • Consider that if a secondary wire is placed in a bifurcation branch, this wire may need to be retracted prior to stent deployment because there is additional risk that the secondary wire may become entrapped between the vessel and the stent.






PRECAUTIONS


Guide wires are delicate instruments and should be handled carefully. Prior to use and when possible during the procedure, inspect the guide wire carefully for bends, kinks, or other damage. Do not use damaged guide wires. Using a damaged guide wire may result in vessel damage and / or inaccurate torque response.


Confirm the compatibility of the guide wire diameter with the interventional device before actual use.


Free movement of the guide wire within the interventional device is an important feature of a steerable guide wire system, because it gives the user valuable tactile information. Test the system for any resistance prior to use. Adjust or replace the hemostatic valve with an adjustable valve if it is found to inhibit guide wire movement.


Never attach the torque device to the modified portion of the proximal end of the extendable guide wire; otherwise, guide wire damage may occur, preventing the ability to attach the DOC™ Guide Wire Extension.


HI-TORQUE™ Guide Wires with Hydrophilic Coating: Avoid abrasion of the hydrophilic coating. Do not withdraw or manipulate the hydrophilic-coated wire through a metal cannula or sharp- edged object.


ADVERSE EVENTS


Potential Adverse Events associated with use of this device may include the following but not limited to perforation, dissection, occlusion, myocardial infarction, embolism and infection.


guide wire.jpg



 

 








0.5409 s.