Share on (60446036951):

FY-TopPette Laboratory Adjustable And Fix Volume Mechanical Pipette
![]() | ![]() | ![]() |
| Single-channel Adjustable & Fixed Volume | Eight-channel Adjustable Volume | Twelve-channel Adjustable Volume |
Features for TopPette
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
Features for multi-channel TopPette
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips
Specification
| Single-channel Adjustable Volume Pipettes | ||||||
| Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | ||
| % | uL | % | uL | |||
| 0.1-2.5μL | 0.05μL | 2.5uL | 2.50% | 0.0625 | 2.00% | 0.05 |
| 1.25uL | 3.00% | 0.0375 | 3.00% | 0.0375 | ||
| 0.25uL | 12.00% | 0.03 | 6.00% | 0.015 | ||
| 0.5-10μL | 0.1uL | 10uL | 1.00% | 0.1 | 0.80% | 0.08 |
| 5uL | 1.50% | 0.075 | 1.50% | 0.075 | ||
| 1uL | 2.50% | 0.025 | 1.50% | 0.015 | ||
| 2-20μL | 0.5uL | 20uL | 0.90% | 0.18 | 0.40% | 0.08 |
| 10uL | 1.20% | 0.12 | 1.00% | 0.1 | ||
| 2uL | 3.00% | 0.06 | 2.00% | 0.04 | ||
| 5-50uL | 0.5uL | 50uL | 0.60% | 0.3 | 0.30% | 0.15 |
| 25uL | 0.90% | 0.225 | 0.60% | 0.15 | ||
| 5uL | 2.00% | 0.1 | 2.00% | 0.1 | ||
| 10-100uL | 1uL | 100uL | 0.80% | 0.8 | 0.15% | 0.15 |
| 50uL | 1.00% | 0.5 | 0.40% | 0.2 | ||
| 10uL | 3.00% | 0.3 | 1.50% | 0.15 | ||
| 20-200uL | 1uL | 200uL | 0.60% | 1.2 | 0.15% | 0.3 |
| 100uL | 0.80% | 0.8 | 0.30% | 0.3 | ||
| 20uL | 3.00% | 0.6 | 1.00% | 0.2 | ||
| 50-200uL | 1uL | 200uL | 0.60% | 1.2 | 0.15% | 0.3 |
| 100uL | 0.80% | 0.8 | 0.30% | 0.3 | ||
| 50uL | 1.00% | 0.5 | 0.40% | 0.2 | ||
| 100-1000uL | 5uL | 1000uL | 0.60% | 6 | 0.20% | 2 |
| 500uL | 0.70% | 3.5 | 0.25% | 1.25 | ||
| 100uL | 2.00% | 2 | 0.70% | 0.7 | ||
| 200-1000uL | 5uL | 1000uL | 0.60% | 6 | 0.20% | 2 |
| 500uL | 0.70% | 3.5 | 0.25% | 1.25 | ||
| 200uL | 0.90% | 1.8 | 0.30% | 0.6 | ||
| 1000-5000uL | 50uL | 5000uL | 0.50% | 25 | 0.15% | 7.5 |
| 2500uL | 0.60% | 15 | 0.30% | 7.5 | ||
| 1000uL | 0.70% | 7 | 0.30% | 3 | ||
| 2-10mL | 0.1mL | 10mL | 0.60% | 60 | 0.20% | 20 |
| 5mL | 1.20% | 60 | 0.30% | 15 | ||
| 2mL | 3.00% | 60 | 0.60% | 12 | ||
8-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
12-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
Fixed Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
5μL | - | 5μL | 1.3% | 0.065 | 1.2% | 0.06 | |||||
10μL | - | 10μL | 0.8% | 0.08 | 0.8% | 0.08 | |||||
20μL | - | 20μL | 0.6% | 0.12 | 0.5% | 0.1 | |||||
25μL | - | 25μL | 0.5% | 0.125 | 0.3% | 0.075 | |||||
50μL | - | 50μL | 0.5% | 0.25 | 0.3% | 0.15 | |||||
100μL | - | 100μL | 0.5% | 0.5 | 0.3% | 0.3 | |||||
200μL | - | 200μL | 0.4% | 0.8 | 0.2% | 0.4 | |||||
250μL | -- | 250μL | 0.4% | 1.0 | 0.2% | 0.5 | |||||
500μL | - | 500μL | 0.3% | 1.5 | 0.2% | 1.0 | |||||
1000μL | - | 1000μL | 0.3% | 3.0 | 0.2% | 2.0 | |||||
2000μL | - | 2000μL | 0.3% | 6.0 | 0.15% | 3.0 | |||||
5000μL | - | 5000μL | 0.3% | 15 | 0.15% | 7.5 | |||||