FY-TopPette Laboratory Adjustable And Fix Volume Mechanical Pipette   















TopPette Single Channel.jpgTopPette - 8 channel.jpg    TopPette - 12 channe.jpg
Single-channel Adjustable & Fixed VolumeEight-channel Adjustable VolumeTwelve-channel Adjustable Volume 

 


Features for TopPette


- Lightweight, ergonomic, low force design


- Digital display clearly reads volume setting


- The pipettes cover volume range of 0.1μl to 10ml


- Easy to calibrate and maintain with tool supplied


- Design helps avoid repetitive strain injuries


- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate


- The low part is available for autoclaving


 


Features for multi-channel TopPette


- 8 and 12 channel pipettes are available for standard 96-well plate


- Dispensing head rotates for optimum pipetting convenience


- Individual piston and tip cone assemblies allow easy repair and maintenance


- Compound material tip cone design allows visual seal verification


- Can be used with universal style pipette tips


 


Specification


















































































































































































































































































Single-channel Adjustable Volume Pipettes
Volume Range  Increment  Test Volume   Maximum permissible systematic error (Inaccuracy)  Maximum permissible random error (Imprecision)
%uL uL 
0.1-2.5μL   0.05μL   2.5uL2.50%0.06252.00%0.05
1.25uL3.00%0.03753.00%0.0375
0.25uL12.00%0.036.00%0.015
0.5-10μL   0.1uL  10uL1.00%0.10.80%0.08
5uL1.50%0.0751.50%0.075
1uL2.50%0.0251.50% 0.015
2-20μL   0.5uL  20uL0.90%0.180.40% 0.08 
10uL1.20%0.121.00%0.1
2uL3.00%0.062.00% 0.04
5-50uL  0.5uL  50uL0.60%0.30.30%0.15
25uL0.90%0.2250.60%0.15
5uL2.00%0.12.00%0.1
10-100uL  1uL  100uL0.80%0.80.15%0.15
50uL1.00%0.50.40%0.2
10uL3.00%0.31.50%0.15
20-200uL 1uL  200uL0.60%1.20.15%0.3
100uL0.80%0.80.30%0.3
20uL3.00%0.61.00%0.2
50-200uL 1uL 200uL0.60%1.20.15%0.3
100uL0.80%0.80.30%0.3
50uL1.00%0.50.40%0.2
100-1000uL 5uL 1000uL0.60%60.20%2
500uL0.70%3.50.25%1.25
100uL2.00%20.70%0.7
200-1000uL 5uL 1000uL0.60%60.20%2
500uL0.70%3.50.25%1.25
200uL0.90%1.80.30%0.6
 1000-5000uL50uL 5000uL0.50%250.15%7.5
2500uL0.60%150.30%7.5
1000uL0.70%70.30%3
2-10mL  0.1mL  10mL0.60%600.20%20
5mL

1.20%


600.30%15
2mL3.00%600.60%12











































































































































































































































































































8-channel Adjustable Volume Pipettes



Volume Range



Increment



Test Volume



Maximum permissible systematic error (Inaccuracy)



Maximum permissible random error (Imprecision)



%



μL



%



μL



0.5-10μL



0.1μL



10μL



1.50%



0.15



1.50%



0.15



5μL



2.50%



0.125



2.50%



0.125



1μL



4.00%



0.04



4.00%



0.04



5-50μL



0.5μL



50μL



1.00%



0.5



0.50%



0.25



25μL



1.50%



0.375



1.00%



0.25



5μL



3.00%



0.15



2.00%



0.1



50-300μL



5μL



300μL



0.70%



2.1



0.25%



0.75



150μL



1.00%



1.5



0.50%



0.75



50μL



1.50%



0.75



0.80%



0.4



12-channel Adjustable Volume Pipettes



Volume Range



Increment



Test Volume



Maximum permissible systematic error (Inaccuracy)



Maximum permissible random error (Imprecision)



%



μL



%



μL



0.5-10μL



0.1μL



10μL



1.50%



0.15



1.50%



0.15



5μL



2.50%



0.125



2.50%



0.125



1μL



4.00%



0.04



4.00%



0.04



5-50μL



0.5μL



50μL



1.00%



0.5



0.50%



0.25



25μL



1.50%



0.375



1.00%



0.25



5μL



3.00%



0.15



2.00%



0.1



50-300μL



5μL



300μL



0.70%



2.1



0.25%



0.75



150μL



1.00%



1.5



0.50%



0.75



50μL



1.50%



0.75



0.80%



0.4



Fixed Volume Pipettes



Volume Range



Increment



Test Volume



Maximum permissible systematic error (Inaccuracy)



Maximum permissible random error (Imprecision)



%



μL



%



μL



5μL



-



5μL



1.3%



0.065



1.2%



0.06



10μL



-



10μL



0.8%



0.08



0.8%



0.08



20μL



-



20μL



0.6%



0.12



0.5%



0.1



25μL



-



25μL



0.5%



0.125



0.3%



0.075



50μL



-



50μL



0.5%



0.25



0.3%



0.15



100μL



-



100μL



0.5%



0.5



0.3%



0.3



200μL



-



200μL



0.4%



0.8



0.2%



0.4



250μL



--



250μL



0.4%



1.0



0.2%



0.5



500μL



-



500μL



0.3%



1.5



0.2%



1.0



1000μL



-



1000μL



0.3%



3.0



0.2%



2.0



2000μL



-



2000μL



0.3%



6.0



0.15%



3.0



5000μL



-



5000μL



0.3%



15



0.15%



7.5



 


 






Contact Us