Share on (60826284260):
Model | Description | Application |
P611 | pH meter | General laboratory pH meter, which can be used in most of the areas related to water quality analysis. |
P612 | Conductivity meter | Used to measure the conductivity value of conventional aqueous solutions, especially in education and scientific research. |
P613 | pH/Conductivity meter | Suitable for high-precision pH/conductivity measurement in scientific research, quality control, biotechnology, fine chemicals,etc. |
P614 | DO meter | Applicable to aquaculture, water source monitoring, sewage treatment and environmental monitoring, etc., also widely used in education and scientific research. |
P615 | pH/DO meter | Suitable for biotechnology, aquaculture, power plants and environmental protection industries, especially in education and research. |
P616 | Conductivity/DO meter | Widely used in biotechnology,aquaculture,power plants and environmental protection industries. |
P618-F | Fluoride ion meter | Test and analysis of the fluoride in drinking water, surface water, sea water, living sewage and industrial wastewater and the fluorine ion in mineral ,organic,and biological materials. |
P618-Na | Sodium ion meter | Suitable for measurement of sodium ion and pH value in natural water,boiler water and industrial drainage,especially for the quality monitoring of high purity water in power plant. |
P618-Cl | Chloride ion meter | Measure chloride ion and pH value in drinking water, food and sewage. |
P618-Ca | Calcium ion meter | Test calcium ion and pH value of water solutions. Widely used in the analysis and research of water quality, biochemistry and biomedicine. |
P618-Br | Bromide ion meter | Test bromide ion and pH value in drinking water, food and sewage. |
P618-NO | Nitrate ion meter | Test nitrate ion and pH value in drinking water, food and sewage. |
P618-K | Potassium ion meter | Measure potassium ion and pH value in drinking water, food and sewage. |
P618-NH | Ammonia Ion Meter | Test ammonia & nitrogen and pH value in drinking water, food and sewage. |
P618-Cu | Copper ion meter | Test copper ion and pH value in food, beverage, electroplating industry and sewage. |
P618-I | Iodide ion meter | Test iodide ion and pH value in drinking water, food and sewage. |
Model | P611 | P612 | P613 | P614 | P615 | P616 | P618 | ||
Standard Accessaries | Electrode holder | √ | √ | √ | √ | √ | √ | √ | |
pH composite electrode | √ | √ | √ | ||||||
Cond. electrode (K=1) | √ | √ | √ | ||||||
DO electrode | √ | √ | √ | ||||||
Ion electrode | √ | ||||||||
Temperature probe | √ | √ | √ | √ | √ | √ | √ | ||
Reference electrode | √ | ||||||||
pH buffer solution | √ | √ | √ | ||||||
DO filling solution | √ | √ | √ | ||||||
Membrane cap | √ | √ | √ | ||||||
Cathode polishing paper | √ | √ | √ | ||||||
Power adapter | √ | √ | √ | √ | √ | √ | √ | ||
Manual | √ | √ | √ | √ | √ | √ | √ | ||
Model | P611 | P612 | P613 | P614 | P615 | P616 | P618 | ||||
Parameters | pH/mV | √ | √ | √ | √ | ||||||
Measuring range | ISE | √ | |||||||||
Resolution | Cond./TDS/Sal./Res. | √ | √ | √ | √ | ||||||
Accuracy | DO | √ | √ | √ | |||||||
Auto calibration | Temp. | √ | √ | √ | √ | √ | √ | √ | |||
pH | Measuring range | (-2.00~19.99)pH | √ | √ | √ | √ | |||||
Resolution | 0.1/0.01 pH | ||||||||||
Accuracy | Meter:±0.01pH; Overall: ±0.02pH | ||||||||||
Auto calibration | 1, 2, 3 point | ||||||||||
Buffer | Europe & USA, NIST, China standards | ||||||||||
mV ORP | Input current | ≤2×10-12 A | |||||||||
Input impedance | ≥1×1012 Ω | ||||||||||
Stability | ±0.01 pH/3h | ||||||||||
pX | Temp. compensation | (0 ~ 100)℃, auto or manual | |||||||||
mV | Measuring range | ±1999.9mV | √ | √ | √ | √ | |||||
Resolution | 1 mV | ||||||||||
Accuracy | ±0.1% FS | ||||||||||
pX | Measuring range | (0.000-14.000) pX | √ | ||||||||
Ion Conc. | Resolution | 0.01 pX | |||||||||
Accuracy | ±0.02 pX | ||||||||||
Temp. compensation | (0 ~ 100)℃, auto or manual | ||||||||||
Cond. | Auto calibration | 1, 2, 3, 4, 5 point | |||||||||
Ion Conc. | Measuring range | (0-19990) | |||||||||
Resolution | 4 significant figures (scientific notation) | ||||||||||
Accuracy | ±0.5% | ||||||||||
Cond. | Measuring range | Conductivity: | √ | √ | |||||||
(0.00~19.99)μS/cm (20.0~199.9)μS/cm | |||||||||||
(200~1999)S/cm | |||||||||||
(2.00~19.99)mS/cm | |||||||||||
(20.0~199.9)mS/cm | |||||||||||
TDS:(0 ~ 100)g/L | |||||||||||
Salinity:(0 ~ 100)ppt | |||||||||||
Resistivity:(0 ~ 100)MΩ·cm | |||||||||||
Resolution | 0.01/0.1/1μS/cm; 0.01/0.1 mS/cm | ||||||||||
Accuracy | ±1.0% FS, Overall:±1.5% FS | ||||||||||
Temp. compensation | (0~100)℃(Auto/manual) | ||||||||||
Electrode constant | 0.1 / 1 / 10 cm-1 | ||||||||||
Reference Temp. | 25℃、20℃、18℃ | ||||||||||
DO | Measuring range | (0 ~ 20.00) mg/L(ppm), (0 ~ 200.0)% | √ | √ | √ | ||||||
Resolution | 0.1/0.01 mg/L(ppm), 1/0.1 % | ||||||||||
Accuracy | Electrode: ±0.10 mg/L, Instrument: ±0.40 mg/L | ||||||||||
Response time | ≤30s (25℃, 90%response) | ||||||||||
Residual current | ≤0.1 mg/L | ||||||||||
Temp compensation | (0 ~ 45)℃, auto | ||||||||||
Salinity compensation | (0 ~ 45) ppt, auto | ||||||||||
Pressure compensation | (66 ~ 200) kPa (manual) | ||||||||||
Auto calibration | Air saturated by water, water saturated by air | ||||||||||
Electrode type | Polarographic | ||||||||||
Temp. | Measuring range | (0 ~ 100)℃ | √ | √ | √ | √ | √ | √ | √ | ||
Resolution | 0.1℃ | ||||||||||
Accuracy | 5 ~ 60℃:±0.4℃ Others: ±0.8℃ | ||||||||||
Other | Display | Large LCD screen | √ | √ | √ | √ | √ | √ | √ | ||
Data storage | 200 sets | ||||||||||
Power | Two AA batteries (1.5V x2) | ||||||||||
Ambient Temp. | 5 ~ 35 ℃ | ||||||||||
Ambient Humidity | ≤85% | ||||||||||
IP grade | IP57 Dustproof and waterproof | ||||||||||
Dimensions/N.W. | (65×120×31)mm/180g | ||||||||||
Shipping Size/G.W. | (460×300×55)mm/1Kg | ||||||||||



















New products from manufacturers at wholesale prices