50ml medical ambulatory syringe pump laboratory syringe pump electrospinning syringe pump

Share on (62313331872):


Price:

Quantity:


Product Overview

Description


Product Description
Detailed Images

10ml medical drug library airless high pressure medical syringe pump

Powerful alarm system with 360 visual light signs, easy for identification.

Double CPU to ensure system security

Manual & automatic bolus, can be programmed according to clinical use;

Automatic anti-bolus to decrease unexpected clinical risk.

Top accuracy in the world: +-2%, flow rate increments by 0.04ml/h.

Easy calibration of syringe and volume etc, For more friendly use.

Drug library contains more than 50 drug names, and can be updated & programmed.

Three infusion modes including Rate, Time, Body-weight.

DPS (Dynamic pressure system)

Model
FC113
Applicable syringes
5ml,10ml,20ml,30ml,50ml/60ml
Infusion Modes
Rate Mode, Time Mode, Body-Weight Mode
Drug Name Display
More than 50,
Can be updated according to customer requirements or clinical use
Accuracy
+-2%
Time Preset
0-99h and 59min
Flow Rate step by
Increments at 0.01ml/h
Flow rate
5ml: 0.01-150ml/h
10ml: 0.01-400ml/h
20ml: 0.01-600ml/h
30ml: 0.01-900ml/h
50ml: 0.01-1800ml/h
60ml: 0.01-1800ml/h
Safetly 
Double CPU to make sure higher standard infusion
Volume limit
0.1-9999.99ml
Purge/Bolus Rate/Top flow rate
5ml: 150ml/h
10ml: 400ml/h
20ml: 600ml/h
30ml: 900ml/h
50ml: 1800ml/h
60ml: 1800ml/h
Audible&visible alarms
syringe dislocation, 
occlusion ,
near end, 
pause overtime,
low battery, 
battery exausted, 
VTBI completion,
syringe empty, 
motor malfunction,
incorrect syringe specification, 
Circult Malfunction,
internal communication malfunction,
master CPU malfunction,
monitoring CPU malfunction.
KVO
Rate≥10ml/h: KVO=3ml/h,
Rate<10ml/h: KVO=1ml/h
Occusion pressure
High: 106.7kpa+-26.7kpa, 
Middle: 66.7Kpa+-13.3Kpa
Low: 40.7kpa+-13.3kpa
Pause overtime alarm
1min 50s-2min
Power & Battery
20VA
AC:100V-240V 50HZ/60HZ
DC:11.1V
Ni-MH battery support from 7 to 10hours
Calssification
Class II, type CF, IPX4
Operation Condition
Temperature: +-5C--+40degree C,
Relative Humidity: 20%-90%
Atmospheric pressure 70Kpa-106Kpa
Optional Function
Event log with at least 2000 history records;DC
Dimensions
140x260x110mm
Weight
2.5kg
Packing & Delivery

0.2577 s.