Model | BK-200 (NEW BK-280) | |||||
Overall Performance | Throughput | 200Tests/Hour | ||||
Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | |||||
Sample & Reagent Unit | Sample Positions | 49 sample positions | ||||
Reagent Positions | 56 reagent positions | |||||
Probewashing | Automaticwashing interior and exterior | |||||
ReagentCooling | Refrigerated tray with independent switch | |||||
Reaction System | Temperature Control | 37±0.1℃, real-time monitor | ||||
Cuvettes | 120 reusable cuvettes,optical length 6mm | |||||
MixerProbe | Independent stirring | |||||
Washing | Automatic cuvettes washing | |||||
STAT function | YES | |||||
Optical System | Light Source | 12V/30W halogen lamp | ||||
Spectrophotometry | Post-spectral spectrophotometry | |||||
Wavelength | 340,405,450,510,546,578,630,700nm | |||||
Absorbance | 0~3.0Abs | |||||
Calibration&QC | Calibration | Linear: K factor, 1-point, 2-point and multipoint linear Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P,Logit-5P | ||||
Quality Control | Real-time QC, Westgard multi rule, Cumulative sum check, Twin Plot(2D) | |||||
Data Management | Software | Windows 7/8/10, LIS system available | ||||
LIS System | Bi-direction, support HL7 protocol | |||||
Interface | LAN port access | |||||
Printer | External, multiple reporting mode available | |||||
Working Conditions | Power Supply | AC220V±10%, 60/50Hz,, 110V±10%, 60Hz, 300W | ||||
Temperature | 10~30℃ | |||||
WaterConsumption | Deionized water:5L/H | |||||
Humidity | 30-80% | |||||
Size & Weight | External Size (W*D*H) | 950*612*510mm | ||||
Net Weight | 75kg | |||||
Package Size(W*D*H) | 1118*728*1151mm | |||||
Gross Weight | 130kg |