Высококачественное оборудование для анестезии yсав 601 для операций в больницах

Сохранить в закладки:


Цена:*
Стоимость в USD:Показать (текущая цена)

Количество:

  • В избранное

Описание и отзывы

Характеристики


high quality anesthesia equipment YSAV 601 for operating hospital emergency


 2017-7-7_01


YSAV601B.jpg


2017-7-7_02


YSAV601 is an economic type of Anesthesia Machine.
The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose onekind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the Customer.
Modularized design and upgrading. Upgrading is available as required by the Customer.


 


Specifications

























































































Part of Anesthesia
Main bodyHigh-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of application adult, child
Gas sourceO2, N2O: 0.27~0.55MPa
Flow meterO2: 0.1~1.0L/min    1.1~10L/min
Flow rate of rapid oxygen supply25~75L/min 
Low oxygen pressure alarmThere will be sound alarm when the oxygen
pressure < 0.2MPa
Vaporizer it has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%. 

Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.


Respiratory circuitworking mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
Respiratory bellowsbellows for adults, bellows for children
tidal volume range:0~1500 ml
Part of Ventilator
Display modeHigh-definition 4.8” LCD screen display
Ventilation modeIPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
Ventilation parametersTidal volume 50~1500ml; Rate 4~80 bpm; I:E 2:1~1:4; Pressure range 1.0~6.0 kPa
Parameters for ventilation monitoringtidal volumeventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform
Security Alarm System
Airway pressure alarmupper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
TV alarmupper limit setting range 0.5~2.0L
low limit setting range 0.02~1.0L
Sustained high-pressure alarmit will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarmit will give sound and light alarm if there is
no tidal volume input in 15s
Power alarm
PowerAC 220V±10%     50Hz±1Hz

 


2017-7-7_032017-7-7_042017-7-7_052017-7-7_06_01


 


0.0093 s.