Сохранить в закладки 1600428051504:
Instructions | JA-600G six channel ECG is such a kind of electrocardiograph,which samples 12 leads ECG signals simultaneously and prints out the ECG waveforms with thermal printing system. Its functions are as follows:recording and displaying ECG waveforms in auto/manual mode;auto-measuring and auto-diagnosing the ECG wave parameters;prompting the state of lead off and paper lack;switching the interface lauguages (Chinese/English);managing the case database. |
Performance | Safety Class: Class I, Type CF applied part. there is defibrillation circuit. Power supply: AC:100V~240V,50/60Hz DC: 7.4V, 3700 mAh rechargeable lithium battery Fuse Specification: 2 pcs φ5×20mm AC time lag; T1.6A/250V(Power Supply:220V) |
Print | Recording way: Thermal printing system Specification of recording paper: 110mm(W)×20m(L) High-speed thermal paper Paper speed: Auto-record:25mm/s,50mm/s,error:±5% Rhythm record:25mm/s,50mm/s,error:±5% Manual_record:5mm/s,6.25mm/s,10mm/s,12.5mm/s,25mm/s,50mm/s,100mm/s,error:±5% |
Input mode | floating and defibrillation protection |
Frequency response | 0.05Hz ~ 150Hz(-3dB~+0.4dB) |
CMRR | >60dB, >100dB(add filter) |
Time | constant:≥ 3.2s |
Patient leak current | <10µA |
Calibration voltage | 1mV |
Sensitivity | 2.5, 5, 10, 20, 40 mm/mV ±5%, standard sensitivity: 10mm/mV±2% |
Noise level | <15µVp-p |
Input circuit current | ≤50nA |
Input impedance | ≥50MΩ |
Sampling accuracy | 12-bit |
Safety classification: | class I, type CF and defibrillation-proof applied part |
Recording mode | thermal printing system |
Auto record | set record according to auto record format and mode, automatically switching leads, measuring and analyzing. |
Rhythm record | set record according to rhythm record format and mode, automatically measuring and analyzing. |
Manual record | set record according to record format, manually switching leads. |
Measurement parameters | HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1) |
Paper size | 80mm(W)×20m(L) |
Power frequency filter | AC 50Hz/60Hz(-20dB) |
Sampling frequency | 1000Hz |
EMG interference filter | 25Hz/35Hz (-3dB) |
Paper speed | Auto: 5, 6.25, 10, 12.5, 25, 50, 100mm/s, ±5% Rhythm: 5, 6.25, 10, 12.5, 25, 50, 100mm/s, ±5% Manual: 5, 6.25, 10, 12.5, 25, 50, 100mm/s, ±5% |