Subject | Range & Accuracy | |
ETCO2 | 0-40mmHg @ ±2%, 41-70mmHg @ ±5%, 71-100mmHg @ ±8%, 101-150mmHg @ ±10%. (0-150mmHg/0-19.7%/0-20kPa) | |
Respiratory rate | 3 - 150 BPM @ ±1 breath | |
Other Specifications | ||
Initialization Time | Capnogram displayed in less than 20 secondsReaching full specifications within 20 minutes. | |
Clibration | No routine user calibration required, the sensor automatic zero when plug inthe adapter No routine user calibration required, The mainstream sensor an airway adapter zero is required when changing to a different airway adapter | |
Compensation | Barometer Pressure 450 - 850 mmHg. User selectable O2 ,N2O,He compensation. | |
Voltage & Power | 5.0 VDC ±5%, power consumption less than 1.3W typical | |
Temp& Humidity | Operating: 0º to 40°, 10 to 90% RH, Storage: -40º to 70ºC, <90% RH | |
Data Interface | Rs232, 19200 baud rate, standard N-8-1. Fully compatible with Respironics CO2 module’s communication protocol | |
Regulatory | Designed to comply YY0601 and ISO21647 | |