Gas Supply | ||
Gas Source | O2,or O2&Air(Option), or O2&N2O(Option); | |
Gas Source Pressure Gauge | O2,or O2&Air(Option), or O2&N2O(Option); | |
O2,280kPa~600kPa/41Psi~87Psi, less than 180L/min;N2O(Option),280kPa~600kPa/41Psi~87Psi,less than180L/min; Air(Option),280kPa~ 600kPa/41Psi~87Psi, less than 180L/min; | ||
Security Valve | release pressure less than 10 kpa; | |
O2 Lost Alarm(Option) | Sound alarm, alimumun cylinder with whistle; | |
N2O cut off valve protection(Option) | when O2 is lost, N2O will be cut off | |
Flow Meter | 2:1~1:6 | |
2 tubes high precision flow meter | ~1L/min,1.1L/min~10L/min; or 0L/min~10L/min(O2&Air or O2&N2O); | |
At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5;O2&N2O Linkage Device(O2&N2O,Option):make insure O2 concentration is not lower than 25%; | ||
O2 Flush | 35L/min~75L/min; | |
One Key Switch | Non-Rebreath; | |
ACGO Port | Non-Rebreath; | |
Selectatec Mounting Bar | single place used for vaporizer, it is easy to replace; | |
Vaporizer | one of Isoflurane/Halothane/Sevoflurane; | |
Respiration System/CO2 Circle Absorber/Rebreath | 0 mL ~ 2000 mL | |
Soda Lime Chamber | 800ml or 1600ml(Option); | |
Pop Off Valve | with one key to close; APL valve(Option): 0.19~6kPa; | |
Airway Pressure Gauge | -20-100cmH2O; |
Anesthesia Ventilator | ||||||||
Drive gas | O2 or Air, Inlet pressure: 280-600 kPa. Max flow≤ 120 L / min; | |||||||
Safety Pressure | System pressure does not exceed 10 KPa; | |||||||
Display | 10.2" TFT, with touch screen; | |||||||
Breath Mode | VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT,SIMV-PRVC(option); | |||||||
Waveform | Pressure,Flow rate,Volume, Loop,PV loop, PF loop,FV loop | |||||||
O2 concentration Monitor | 15%-100%(option); | |||||||
Electric PEEP | OFF, 3 ~ 30 cmH2O; | |||||||
Battery Type | Li-ion,work time more than 3 hours; | |||||||
Parameters setting range | ||||||||
Tidal volume | 50 - 1500ml; | |||||||
option | 10 - 1500ml; | |||||||
Respiratory rate | 1 ~ 100 bpm; | |||||||
Inspiratorytime | 0.1~10.0s; | |||||||
Respiratory rate | 0 ~ 100 bpm; | |||||||
Percentage of inspiratory pause | 0 to 60%; | |||||||
PEEP | OFF, 3 ~ 30 cmH2O; | |||||||
Pressure Support | 0 ~ 70 cmH2O; | |||||||
Pressure Control | 5 ~ 70 cmH2O; | |||||||
Flow trigger | 0.5 ~ 20L/min; | |||||||
Pressure Trigger | 0~ 20 cmH2O; | |||||||
PSV Insp Termination Level | 5 ~ 80%; | |||||||
Monitored parameters | ||||||||
Inspiratory tidal volume | 0 ~ 3000 mL; | |||||||
Expiratory tidal volume | 0 ~ 3000 mL; | |||||||
Spontaneous minute ventilation | 0 ~ 60 L / min; | |||||||
Minute ventilation | 0 ~ 60 L / min; | |||||||
Respiratory ration | 4:1 to 1:10; | |||||||
Spontaneous breathing frequency | 0 ~ 100 bpm; | |||||||
Compliance | 0 ~ 300 mL/cmH2O; | |||||||
Respiratory ratio | 30:1 to 1:150; | |||||||
Peak airway pressure | 0 ~ 100 cmH2O; | |||||||
Mean airway pressure | 0 ~ 100 cmH2O; | |||||||
PEEP | 0 ~ 100 cmH2O; | |||||||
Inspiratory plateau pressure | 0 ~ 100 cmH2O; | |||||||
FiO2(option) | FiO2(option) | |||||||
Tidal volume high/low alarm; Minute ventilation high/low alarm; Respiratory rate high/low alarm; Airway pressure high/low alarm; Continuous Positive Airway Pressure/Negative pressure alarm; Apnea alarm; Oxygen deficiency alarm; AC power failure alarm ,Low battery alarm. | ||||||||
Anesthesia machine, one unit X60 Anesthesia vaporizer, one set adult disposable breath circuit, one set pediatric disposable breath circuit, 0.5,1.2,3L airbag, O2 gas source pipeline | ||||||||
Dimension&Weight | Trolley Stand: 55cm*56cm*153cm,50kg; |