Полностью автоматический набор для экстракции ткани геномной ДНК магнитные шарики рабочая станция

Сохранить в закладки 1601250359608:


Цена:6 140 310,40 ₽*

Количество:

  • В избранное

Описание и отзывы

Характеристики



If you need a discount distribution price, please contact us.


Nucleic Acid Purification System Auto-Pure 2400




Product Description


Auto-Pure series is a high-throughput automatic nucleic acid extraction equipment, based on the magnetic bead method to extract nucleic acid. During the purification process, the steps of pipetting, lysis, washing, elution, and PCR system set-up can be done automatically in the workstation. One-stop operation, without manual intervention, can extract 24×10 mL samples at the same time, which is more efficient and convenient than manual and semi-automatic extraction. At the same time, standardized and automated operating steps avoid human error, the experimental results are more reliable and repeatable.
In addition, the full-coverage UV sterilization system and HEPA negative pressure filter system effectively prevent cross-contamination. As an efficient intelligent workstation, Auto-Pure 2400 series can bring you a better product experience.






Features


•Large volume extraction to meet special needs
The maximum reaction system of the sample library is 10 mL, which can meet the needs of cfDNA extraction
The elution hole is eluted in a small volume to meet the nucleic acid concentration requirements

•High-precision liquid handling
Dual-channel pipetting design, 1×50-5000 μL, suitable for 1 mL / 5 mL pipettor and 1×1-200 μL, suitable for 200 uL pipettor
The pipettor's pressure detection technology can not only detect the height of the liquid level, but also empty suction, residual liquid volume and blockage detection
Detailed pipetting parameter settings for different liquids can effectively improve the accuracy

•Methylation transforms special functional modules
Module temperature control range 4 °C ~ 105 °C
In order to meet the methylation requirements, 2400M adds a temperature control module to realize multi-stage temperature change
Thermo lid design can prevent reagent condensation from affecting the reagent concentration in the tube, and has the technology to prevent cross-contamination

•Easy-to-learn operating software
Graphical user interface, program editing can be completed without complicated operations
Built-in comprehensive libraries of consumable carriers, and liquids to reduce customer program matching time
Users can simply select according to their needs to complete the whole process of liquid handling,extraction, and PCR / qPCR set-up

•Open and reliable nucleic acid extraction system
Open software design, suitable for most magnetic bead extraction reagents on the market
A variety of phosphorus absorption methods to effectively solve the problem of phosphorus beads residue



Technical date


Model
Auto-Pure 2400
Function
Automatic pipette samples and reagents, extract nucleic acid and set-up PCR systems
Information management
Support external USB scanner, barcode scanning function, automatic scanning function
Pipetting principle
Air displacement pipetting technology
Pipettor
One 5000 μL pipettor, one 200 μL pipettor
Pipetting range (μL)
1-5000 μL
Tips
Allsheng tips, and other tips compatible with this instrument
Liquid detection
Pressure detection: liquid level detection, tip clogging detection and residual liquid detection
Samples per run
1-24
Extraction processing volume
50-10000 μL
Extraction uniformity
CV≤5 %
Single sample consumable
Support
Extraction temperature control
Elution, lysis temperature control: RT~ 120 ℃
UV sterilization
Separate UV lamp: extraction area, pipetting area and waste container area
Filter
Equipped with three HEPA filters
Instrument port
USB port
Power supply
220 V, 8 A, 50/60 Hz
Power
Standby 30 W, maximum 600 W
Noise
≤60 dB
Noise
130 kg