Bioer FQD-48A Высококачественная система PCR в реальном времени экстракция нуклеиновой кислоты DNA/RNA для обнаружения
2 148 953,60 ₽ - 7 162 994,96 ₽
Сохранить в закладки 1601273478907:
Basic parameter | ||
Model | Leap-Pure S24 / H24 | |
Working principle | Magnetic bead method, magnetic rod type, pipetting type / tip blowing and aspirating mixing | |
Sample | FFPE, whole blood, plasma, stool, sputum, etc | |
Consumable | Leap-Pure S24: 2 mL / 8 mL / 15 mL cartridge Leap-Pure H24: 20 mL cartridge | |
Throughput | 1~24 | |
Sample processing system | Leap-Pure S24: 1 mL~15 mL Leap-Pure H24: 1 mL~20 mL | |
Elution volume | 20~200 μL [the minimum processing volume of 20 μL (FFPE block / 1 mL cfDNA)] | |
Stability | CV ≤5 % | |
Software operation | 10-inch color touch screen, windows system | |
Account management | User authorization management, programs and instruments can be customized | |
Instrument port | USB port, LAN port | |
Information management | Support external code scanner | |
Pollution control | Built-in HPEA, UV sterilization | |
Power supply | AC 100~240 V, 50 Hz, rated power 700 W | |
Dimension (W×D×H) | 740 mm×640 mm×590 mm | |
Weight (kg) | 85 kg | |
Extraction temp. control | ||
Average heating rate | 40 °C~100 °C ≥0.2 °C/s (at 25 °C) | |
Temp. control range | 37 °C~120 °C | |
Temp. accuracy | ±1 °C @ 100 °C (at 25 °C) | |
Temp. uniformity | ≤3 °C @ 100 °C (at 25 °C) | |
Temp. stability | ≤±0.5 °C @ 100 °C (at 25 °C) | |
Pipetting block | ||
Pipetting range | 20~1000 μL | |
Pipetting precision | 20 μL: ≤5 % 100 μL: ≤1.5 % 500 μL: ≤1 % 1000 μL: ≤1 % | |
Pipetting accuracy | 20 μL: ±8 % 100 μL: ±2 % 500 μL: ±1 % 1000 μL: ±1 % |
Code | Product Description |
AS-17250-00 | Leap-Pure S24 nucleic acid purification system |
AS-17260-00 | Leap-Pure R16 nucleic acid purification system |
AS-17270-00 | Leap-Pure H24 nucleic acid purification system |