Meanwell RPS 120 медицинский источник питания 24V12V27V импульсный 15V48V низкий ток утечки MPS высокий иммунитет C PCB

Сохранить в закладки 1601350541159:


Цена:*

Количество:

  • В избранное

Описание и отзывы

Характеристики






























































































































































































































































































modelRPS-120-12(S)RPS-120-15(S)RPS-120-24(S)RPS-120-27(S)RPS-120-48(S)
outputvolts d.c12V15V24V27V48V
current10CFM10A8A5A4.5A2.5A
natural air cooling7.0A5.6A3.5A3.15A1.75A
power rating10CFM120W120W120W121.5W120W
natural air cooling84W84W84W85W84W
Waves and noise (maximum)100mVp-p120mVp-p150mVp-p150mVp-p150mVp-p
Voltage adjustment range11.4~12.6V14.3~15.8V22.8~25.2V25.6~28.4V45.6~50.4V
voltage accuracy±2.0%±2.0%±1.0%±1.0%±1.0%
Linear adjustment rate±0.5%±0.5%±0.5%±0.5%±0.5%
load regulation±1.0%±1.0%±1.0%±1.0%±1.0%
Start-up, rise time500ms, 30ms / 230VAC 500ms, 30ms / 115VAC (when full load)
Hold time (Typ.)50ms / 230VAC 10ms / 115VAC (when fully loaded)
importvoltage range80~264VAC     113~370VDC
frequency range47~63Hz
productiveness (Typ.)89%89%90%90%91%
AC current (Typ.)2.1A/115VAC    1.2A/230VAC
Inrush current (Typ.)Cold start: 30A / 115VAC 60A / 230VAC
leakage currentFor the floor drain current <150 μ A / 264VAC, and for the contact drain current <80 μ A / 264VAC
protectexcess load115~150% of the rated output power
Protection mode: hiccup mode, which can be automatically restored after the abnormal load condition is removed
overvoltage13.2~15.6V16.5~19.5V26.4~31.2V29.7~35V52.8~62.4V
Protection mode: turn off the output, and restore the normal output after the power supply restart
excess temperatureProtection mode: turn off the output, and restore the normal output after the power supply restart
functionFan-assisted power supply12V@ 0.5A Provide power to the fan; if the main output is 40% rated load (10 CFM), the voltage accuracy is-15%~ + 15%
environmentworking temperature-30~+70℃
Working humidityFrom 20 to 90% RH, with no condensation
Storage temperature, and humidity-40~+85℃,10~95%RH
temperature coefficient±0.03%/℃(0~50℃)
ibration resistance10-500 Hz, 2G 10 minutes / cycle, 60 minutes each X, Y and Z
Operate altitude4000 Meters
Safety and electromagnetic compatibilitysafe codeIEC60601-1:2005 + A1 + A2, TUVBS EN / EN 60601-1:2006 + A1 + A12 + A2, ANSI / AAMI ES60601-1:2005 + A2 CAN / CSAC22.2 No.60601-1:2014 + A 2, EACTPTC004 certified; design refer to BSEN / EN60335-1 (as required)
Isolation levelPrimary-secondary: 2 xMOPP, primary-ground: 1 xMOPP, secondary-ground: 1 xMOPP
withstand voltageI/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
Insulation impedanceI/P-O/P:100M Ohms/500VDC/25℃/70%RH
Electromagnetic compatible emissionParameterStandardTestLevel/Note
Conducted emissionBS EN/EN55011 (CISPR11)Class B
Radiated emissionBS EN/EN55011 (CISPR11)Class B
Harmonic currentBS EN/EN61000-3-2Class A
Voltage flickerBS EN/EN61000-3-3-----
EMC immunityBS EN/EN55035,BS EN/EN60601-1-2
ParameterStandardTestLevel/Note
ESDBS EN/EN61000-4-2Level 4,15KV air;Level 4,8KV contact
RF field susceptibilityBS EN/EN61000-4-3Level 3,10V/m(80MHz~2.7GHz)Table 9,9~28V/m(385MHz~5.78GHz)
EFT burstsBS EN/EN61000-4-4Level 3,2KV
Surge susceptibilityBS EN/EN61000-4-5Level4,4KV/Line-FG;2KV/Line-Line
Conducted susceptibilityBS EN/EN61000-4-6Level 3,10V
Magnetic field immunityBS EN/EN61000-4-8Level 4,30A/m
Voltage dip,interruptionBS EN/EN61000-4-11100%dip 1 periods,30%dip 25 periods,100%interruptions 250 periods
elseMTBF3732.9K hrs min.Telcordia SR-332(Bellcore);653.6K hrs min.MIL-HDBK-217F (25℃)
sizePCB: 101.6 * 50.8 * 29mm or 4 "* 2" * 1.14 "inch; enclosure type: 103.4 * 62 * 40mm or 4.07" * 2.44 "* 1.57" inch
packPCB:0.15Kg;72pcs/11.8Kg/0.84CUFT; Shell type: 0.24Kg;60pcs/15.4Kg/1.06CUFT

 


 


 


 


RPS-120   1.pngRPS-120   2.png


Похожие товары

Новые поступления

Новинки товаров от производителей по оптовым ценам