Basic description | ||||
Parameter | Description | |||
Display | 9 inch TFT, with touch screen. | |||
Drive gas | Gas type | O2 or Air | ||
Inlet pressure | 280-600 kPa. | |||
Max flow | ≤ 120 L / min. | |||
Operating Mode | VCV,PCV,SPONT,Manual,Demo | |||
Waveform | Waveform: Pressure,Flow-rate,Volume Loop:PV loop, PF loop,FV loop | |||
Safety Pressure | System pressure does not exceed 6 KPa. | |||
Battery Type | Li-ion,work time more than 2 hours |
Parameters setting range | ||
Parameter | description | |
Tidal volume | Range: 20 - 1500ml; Increment: 1 mL; | |
Respiratory rate | Range: 1 ~ 100 bpm; | |
Inspiratory time | Range: 0.1 ~ 10.0 s | |
Respiratory ratio | Range: 9.9:1 to 1:9.9 | |
Percentage of inspiratory pause | Range: 0 to 50% | |
PEEP | Range: OFF, 3 ~ 20 cmH2O | |
Pressure Support | Range: 0 ~ 60 cmH2O | |
Pressure Control | Range: 5 ~ 60 cmH2O | |
Flow trigger | Range: 0.5 ~ 20L/min | |
Pressure Trigger | Range: -1 ~ -20 cmH2O | |
PSV Insp Termination Level | 25%. |
Monitored parameters | ||
Parameter | Description | |
Inspiratory tidal volume | Range: 0 ~ 2500 mL | |
Expiratory tidal volume | Range: 0 ~ 2500 mL | |
Minute ventilation | Range: 0 ~ 60 L / min | |
Spontaneous minute ventilation | Range: 0 ~ 60 L / min | |
Respiratory rate | Range: 0 ~ 100 bpm | |
Spontaneous breathing frequency | Range: 0 ~ 100 bpm | |
Respiratory ratio | Range: 9.9:1 to 1:9.9 | |
Peak airway pressure | Range: 0 ~ 100 cmH2O | |
Mean airway pressure | Range: 0 ~ 100 cmH2O | |
PEEP | Range: 0 ~ 100 cmH2O | |
Inspiratory plateau pressure | Range: 0 ~ 100 cmH2O | |
FiO2(optional) | Range: 15 to 100% |
Alarm settings | ||||
Alarm type | Range | |||
Tidal volume | High | 0 ~ 2000 mL, OFF | ||
Low | 0 ~ 1500 mL | |||
Minute ventilation | High | 1 ~ 99 L | ||
Low | 0 ~ 98 L | |||
Respiratory rate | High | 1 ~ 100bpm | ||
Low | 0 ~ 99 bpm | |||
FiO2(optional) | High | 19 ~ 100% | ||
Low | 18 ~ 99% | |||
Airway pressure | High | 6 ~ 99 cmH2O | ||
Low | 0 ~ 9 8 cmH2O | |||
FiCO2 (optional) | High | 0.1 to 13.3%; | ||
Low | 0 to 13.2% | |||
Continuous Positive Airway Pressure | Airway pressure greater than (PEEP +15) cmH2O, 15 s sustained when an alarm is triggered. | |||
Negative pressure alarm | Airway pressure is less than (-10) cmH 2 O. | |||
Apnea alarm | Setting time is 10 ~ 60 s, in increments of 1 s. | |||
Gas deficiency alarm | Gas pressure is less than 280 kPa. | |||
AC power failure alarm | The mains breaks down or the power cord disconnects | |||
Low battery alarm | Alarm battery time is 20 min. | |||
Alarm battery is exhausted | Alarm battery time is 10 min. | |||
Silent Alarm | ≤ 120 s | |||
O2 sensor failure | FIO2 <15% Vol; |