ZYCGF Conventional Analysis Reverse Osmosis Water Purification System
The model ZYCGF reverse osmosis water purification system which uses the technology of reverse osmosis and polishing resin to purify the dissolved salt and particlar from feed water and make the oulet water ultra pure water (Deionized water).
The ultra pure water (DI water) is usually used for laboratory test .
| Model | Bench type | Floor type | Bench type | Floor type |
| ZYCGF-I-5/10/20T | ZYCGF-I- 20/40/60/100L | ZYCGF-II-5/10/20T | ZYCGF-II- 20/40/60/100L | |
| Inlet water | Total dissolved solids contect TDS<200ppm, hydraulic pressure1.0-5.0kg/cm2,Optimum 25℃ operating temperature | |||
| Water yield | 5/10/20L/H | 20/40/60/100L/H | 5/10/20L/H | 20/40/60/100L/H |
| Effluent velocity | 1.5-1.8L/MIN (When Water storage tank working) | |||
| RO water quality (μs/cm) | Condutivity <2-10μs/ cm | Condutivity <2-10μs/ cm (Online monitoring) | ||
UP water Effluent quality (MΩ.cm) | Resistivity 17-18.25MΩ.cm (Online monitoring) | Resistivity 18.25MΩ.cm (Online monitoring) | ||
| Size (mm) | 380*500*590 | 380×580×950 | 380*500*590 | 380×580×950 |
| Weight (kg) | 35kg | 70kg | 35kg | 70kg |
| Power (w) | 30W-50W | 50W-100W | 30W-50W | 50W-100W |
Application |
sterilization pot; detection; conventional qualitative and quantitative analysis; | 1, microbial medium chromatography (HPLC); chromatography (HPLC) and other instrumental analysis of water. | ||
External pressure tank | A.standard configuration 3.8G | A. standard configuration 11 | A. standard configuration 3.8G | A. standard configuration 11 |
| Ordering instructions: | ||||
| • Optional configuration:UV digestion or UF ultrafiltration column, namely III or IV machine | ||||
• This water production is according to the rated water of RO Membrane with water temperature 25℃ ,When the water temperature is reduced by 1 DEG, the water yield of the RO film is decreased by about 3%, when the water temperature is below the temperature of 5 DEG , the RO film will stop producing water, and when the user selects the order, the water temperature in the area should be taken into account. | ||||
• The area with poor Source water quality (TDS>200ppm) ,RO pure water conductivity may be reduce to more than 10μs/cm; the users require higher water quality proposed matching ZYROEU (reverse osmosis) or pre watersoftener system (multi medium filter + activated carbon filter + water softener). | ||||



The technology process is divided into source water pretreatment system,reverse osmosis system and ultra-pure system.
1. Source water pretreatment system: PP fiber filter + AC filter + Softener
2. Reverse osmosis system: It has high removal rate more than 99% for high valent ions, colloids, bacteria and organic matter (including endotoxin) with molecular weight greater than 300Dalton.
3. Ultra pure system: Ion exchange resin using H + exchange cation out the water in the filter ,using OH- exchange anions out the water , the replaced H + and OH- combined to produce H2O, ultra-purified column filling the United States Dow UPW-450 Ultra-pure water for nuclear grade resin, the water resistance of up to 18.25 MΩ.cm.


Ultra Pure Water System Equals to Lab Aqua Distilled Water ZYCGF

aqua distilled water
Why choose us :
1. Professional manufacturer : More than 10 years experience in prodcuing pure water treatment system .
2. Strict time control : Ensure delivery time.
3. Factory strength: The factory area is more than 2000m3, more than 100 workers.
4. After-sales service : 24 Hours online for service.
5. Customized is available. Sample order is available.