Nephelometry | Linear range | 100~4000 NTU Linear correlation coefficient≥0.99 |
RSD | RSD≤8% | |
Stability | The turbidity variation≤10% | |
Accuracy | Relative deviation≤10% | |
Fluorescence | Linear range | 100~1600AU, Linear correlation coefficient r≥0.99 |
RSD | RSD≤8% | |
Stability | Fluorescence variation≤15% | |
Accuracy | Relative deviation≤10% | |
Sensitivity | Detection limit is not more than 2.5μmol/L | |
Absorbency | Linear range | See attached table 1 |
RSD | RSD≤1% | |
Stability | The absorbency deviation is not more than 0.005 at 340nm position within 20min | |
| Inter-channel RSD | RSD≤5% |
Test temp. accuracy | 37℃±0.5℃ | |
Wavelength accuracy | The wavelength deviation of 340nm,550nm,650nm is±3nm,and half width is no more than 12nm | |
Supply voltage | 220V/50Hz | |
Input AC power | <100VA | |
Size | 410mm×280mm×355mm | |
Weight | 5kg | |
Security classification | Class I for no application part |
Attached table 1 Linear range of absorbency
Linear range (A) | Deviation(%) |
0.2~≤0.5 | -5-+5 |
>0.5~≤1.0 | -4-+4 |
>1.0~≤1.8 | -2-+2 |
25 Tests/Kit
40 kits/Carton
Hipro is a high-tech enterprise which originate in Silicon Valley, USA. We delicate ourselves to IVD for last 19 years and focus on R&D, manufacturing, marketing, and relevant services of Point-of Care products. Hipro has a world-class high capacity manufacturing facility (10000 sq.m.) with 1500 sq.m. current GMP plants and strict quality control system. The quality system is ISO 13485 and CE.