No. | Item | Specified | S50 Offered | ||
1 | General Requirement | ||||
◆ | Released at least 1 years | released at least at 2015 year | |||
◆ | Operation System | Linux Platform | |||
◆ | Size | maximum: L×W×H(mm): 987×586×1320 | |||
Multiple Beam Processing | required | ||||
Applications | have a wide range of applications includes: abdominal, OB/Gyn, cardiology, urology, small parts, vascular, orthopedic, anesthesia and MSK. | ||||
◆ | Monitor | no less than 21.5 inch, 1920×1080 high resolution LED monitor, widescreen,anti-flickering, can be vertically and horizontally swiveled | |||
◆ | Touch screen | required, no less than 12.1 inch, and select the Probe with the probe name and icon, and show different presets with the body Anotomy icon | |||
﹡ | Touch screen control | The position of the parameters can be controled and adjusted in the touch screen | |||
﹡ | Keyboard | PC Keyboard on the control panel, with at least 2 user-defined keys | |||
A/D bit control | at least 14 A/D bit | ||||
◆ | GPU | required | |||
﹡ | Digital channel | at least 36864 digital channel | |||
Operation Mode | B | ||||
Dual B | |||||
Quad B | |||||
Pluse inversion Harmonic Imaging ( PHI ) | |||||
Trapezoid Imaging | |||||
Real-time Panoramic Imaging(B mode ) | |||||
Real-time Panoramic Imaging(Color mode and DPI mode) | |||||
Panoramic support in Linear and convex. Velocity indicator and image rotate. | |||||
M | |||||
Color M | |||||
Anatomic M (at least 3 sample lines) | |||||
Color Doppler | |||||
Directional PDI | |||||
TDI | |||||
Power Doppler Imaging | |||||
PW with HPRF, CW | |||||
Dual-Live | |||||
Duplex: B and Doppler/M, can be defined in new preset | |||||
Triplex: B, Color Flow, and PW/CW Doppler,can be defined in new preset | |||||
3D/4D Imaging with S-Live function(light source at least 8 position) | |||||
Scr-Zoom | |||||
Compound Imaging (0,1,2) | |||||
u-scan | |||||
Needle enhancement ( Vis-needle ) | |||||
Contrast Imaging | |||||
C-xlasto( elastography imaging) | |||||
﹡ | Transducer Sockets | at least 5 transducer sockets, 4 active transdcuer connectors, which can support all transducers. Should have a special transdcuer connector for pencil probe. | |||
◆ | Transducers | 1.can supply with more than 9 transducer choices, covering linear, convex, endocavity, phased array and 4D transducers. Convex, Linear, probe should be high element probes which is no less than 128 elements(Convex), 256 elements(Linear). 2.can supply with single cystal probe, covering convex and phased array transducers. | |||
Scanning Methods | Electronic Convex Secter Electronic Linear Secter Electronic Phased Array Sector | ||||
ECG | support ECG function | ||||
﹡ | Clipboard Function | Capture and reviewthe archived images, support output the images and viedo to USB and DVD in scanning interface | |||
Image Optimization 1 | Images can be optimized by one button in B mode | ||||
Image Optimization 2 | Images can be optimized by one button in PW/CW mode | ||||
◆ | Image Optimization 3 | Images can be optimized by one button in Color mode (Automatically dectect the Color flow with ROI) | |||
Screen Saver | users can choose 0-99minutes | ||||
TGC | 8 segments control | ||||
Zoom | The zoom value can be shown.The minimum zoom ratio can low to 0.8. The max zoom ratio can up to 10. Support PIP. | ||||
Auto foucs | support B mode and color mode | ||||
Biposy | support 2 needle guide line and 1 gun guide line | ||||
﹡ | Warranty | spare parts should be provided with system for 18 months | |||
2 | Scanning Parameters | ||||
B Mode | |||||
◆ | Focus | focus number can up to 12 points, and the intra-focus distance can be adjustable | |||
TGC | 8 slider controls | ||||
Dynamic Range | at least 20-280dB(Probe dependent) can be adjusted by 20dB/step. | ||||
Gain | 1~255, adjustable | ||||
﹡ | Depth | 40.0cm Maximum | |||
Chroma | Maximum 13 selectable | ||||
Frequency | 5 bands | ||||
Persistence | 0-60 | ||||
Grayscale curve | 7 types | ||||
Adaptive image fusion | 15 types | ||||
﹡ | Line Density | can choose the lower line density for higher frame rate, and the maximum frame rate can up to 1493 f/s, and line density can be adjusted by 3 levels: low/med/high | |||
◆ | u-scan | Speckle Reduction Imaging,adjustable | |||
Compound Imaging | required | ||||
Live Panoramic imaging | required | ||||
Trapezoidal imaging | is availiable for linear array transducer | ||||
Biopsy guide function | required,and the biopsy line adjustable. | ||||
Imaging width and position | adjustable | ||||
Tissue acousitc | at least 1400-1700m/s selectable | ||||
Image Optimization by one button | required | ||||
B-steer | required,at least the Linear probe | ||||
Color Flow Mode/TDI | |||||
◆ | Live Panoramic imaging | Required | |||
◆ | Frame Rate | Minimum 211 f/s | |||
Pulse Repetition Frequency | 0.5~12KHz | ||||
Auto focus | Required | ||||
Steer Angle | up to 20 degrees | ||||
Wall filter | 25-750 | ||||
Baseline | ±15 steps, adjustable | ||||
Persistence: | at least 0-80 | ||||
Image orientation | left/right, up/down | ||||
Flow Invert | required | ||||
Color /Direction energy | at least 14 kinds,selectable | ||||
Line density | can be adjusted by 3 levels:low/medium/high | ||||
B and Color Flow Simultaneous Realtime Display | required | ||||
◆ | Color Flow | available in the frozen mode | |||
◆ | Flow velocity | available in the frozen mode | |||
◆ | Real Time Flow Volume Analysis | required | |||
M Mode | |||||
Sweep Speed | adjustable, at least 6 levels | ||||
Video Inversion Function | required | ||||
Chroma | at least 5 types | ||||
Display Format | at least 6 types | ||||
◆ | Steer M | 3 lines and Frame rate can be displayed | |||
Spectral Doppler | |||||
PRF | PW : 1~16KHz CW: 1-48KHz | ||||
Maximum Velocity Range | PW: 0.0004~18 m/s CW: 0.0013~63 m/s | ||||
Baseline | 17 steps | ||||
Frequency | 5 bands | ||||
﹡ | Angle Correction | 0~72 degree, 2 degree per step or less | |||
Auto Angle Correction | Auto 60 degree of Angle correction by one botton | ||||
Sample Volume in PW Doppler | 0.7~21 mm, adjustable | ||||
Sweep Speed | adjustable, at least 4 levels | ||||
Chroma | 5 types | ||||
Triplex | required | ||||
Video Inversion Function | required | ||||
Spectrum Inversion | required | ||||
﹡ | Real Time Auto-Trace | required | |||
Wall Filter | 25-750, adjustable | ||||
◆ | One-key Auto Optimization | At least can adjust the baseline, PRF,and correct angle | |||
Steer angle | 0,±16, ±20 | ||||
Display Format | 6 | ||||
Deflection Angle | 5 levels, adjustable | ||||
3D/4D Mode | |||||
Functions | include Trace Cut, Undo Cut,Clip Plane, Multi-slice, Cine Review | ||||
Display Mode | Dual, Quad, Full 4D, 3D | ||||
Adjust Slice | X Rotation, Y Rotation, Z Rotation | ||||
Opacity Threshold | 0-255 | ||||
◆ | Render Mode | Surface, Grad.light , Skeleton, S-live ,S-depth, Transp.Min, X-Ray,(At lease has 7 types of Render Mode) can render in real time | |||
Auto Rotate | 0°,45°, 90°, 180° , 270°and 360°, selectable | ||||
Zoom In/Ou | adjustable | ||||
3D Chroma | 0-14 adjustable | ||||
Color Map | 1-13 adjustabe | ||||
Transparency | 0-100 adjustable | ||||
Contrast | 0-100 adjustable | ||||
Brightness | 0-100 adjustable | ||||
Smoothness | 0-30 adjustable | ||||
Multi-Slice | required, Slice number: 1*2, 2*2, 3*4, 3*3, 4*4, 5*5 | ||||
Slice Spacing | 0.5-2.0, adjustable | ||||
Measurement | Distance, Area, Volume | ||||
◆ | AVC foilicle | Left Ovary, Right Ovary | |||
Storage | 3D Image/Cine/Volume Storage | ||||
Cutting Line Curvature and position | adjustable | ||||
Image Quality of 4D | Low/Medium/High, selectable | ||||
Rescan | required | ||||
4D Gain | adjustable | ||||
Frame Rate | 40 f/s | ||||
Print | required | ||||
3 | Transducer | ||||
﹡ | Transducer Warranty | transducers are manufactured by supplier, and the warranty should last 18 months | |||
﹡ | Transducer Frequency | broadband (1.0~17MHz) with multiple 5 frequencies range in B Mode/Color Mode/PIH Mode/PW mode/CW/Mode | |||
◆ | Transducer Elements | at least 192, and can optional 256 | |||
﹡ | Linear Transducer | frequency range: 3~17MHz | |||
◆ | Convex Transducer | frequency range: 1~15MH , support single cystal convex transducer | |||
◆ | Endocavity transducer | frequency range: 3~15MHz scanning angle: 200° | |||
◆ | Temperature-detection Technology | the temperature of endocavity transducer can be displayed | |||
◆ | Phased Array | frequency range: 1~7MHz , support single cystal phased array transducer | |||
Biopsy Guide | Required | ||||
4 | Image Storage | ||||
Cine Loop | available for all mode | ||||
﹡ | Cine Loop Memory | 1000 frames in B mode, 8000 frames in Contrast mode | |||
Image Format | JPG, BMP, TIF | ||||
Clipboard function | is avilibale in frozen B mode | ||||
Clip Format | WMV, AVI | ||||
One Key Save | required | ||||
Doppler cine playback | speed is adjustable, sound can be played back | ||||
Print&Store Region | Full Screen, Image&Result Area, US Image | ||||
Image Share Service | required | ||||
Store To UsbDisk | required | ||||
Storage Capability | Hard Disk can up to 500G/1T | ||||
Report Tables | Abdominal, OB/Gyn, Cardiology, Urology, Small Parts, Vascular,IMT | ||||
Report Format | TXT, PDF | ||||
◆ | Report Templet | at least 6 images can be displayed in report | |||
5 | Measurements | ||||
Measurement Packages | Abdominal, OB/Gyn, Cardiology, Urology, Small Parts, Vascular, Orthopedic, and measurement packages can be chosen for Icon Driven or All Packages | ||||
Measurement in B Mode | Distance, Angle, Ratio(Distance,Area), Area/Circumference(Ellipse,Trace),Volume(3 axis/2 Axis) | ||||
Measurement in M Mode | Distance, Time, Slope, HR | ||||
Measurement in PW Mode | Flow Velocity, Acceleration, Time, Heart Rate, and has automatic measurement functions | ||||
◆ | Quick Measurements | Measurement can be select by one button | |||
﹡ | IMT Measurement | support IMT measurement, and the result can be displayed in report | |||
◆ | Auto NT | support Automatic NT measurement | |||
◆ | Auto EF | support Automatic EF measurement | |||
Obstetrics Packages | |||||
Measurement items | 28 types, including GS, CRL, BPM, HC, AC and AFI. | ||||
Formula | users can define the formula by themselves | ||||
◆ | Fetal Growth Chart | systems can display fetal growth in a graph chart according to some measurement items and their OB tables. Up to 4 charts | |||
Cardiac Packages | |||||
B Mode | Single-plane Ellipse, Bi-plane Ellipse, Bullit, Simpson | ||||
M Mode | Cube, Teichholz, Gibson | ||||
﹡ | PW Mode | Real Time Automatic Measurement | |||
6 | Annotations | ||||
Comments | no less than 130 comments in dictionary, and users also can define it by themselves | ||||
Arrow | Orientation adjustable | ||||
﹡ | Body Mark | up to 109 types,can be classified by specific exams | |||
7 | Input and Output | ||||
◆ | DVD-R/W | Required | |||
Footswitch | Required | ||||
S-video output | Required | ||||
﹡ | USB Port | 4 ports | |||
VGA/HDMI output | Required | ||||
Printer interface | Video printer | ||||
Ethernet port for DICOM | Required | ||||
8 | Preset Function | ||||
Preset Function | Required, can be customized | ||||
The sequency of presets | Users can arrange the presets | ||||
import or export presets | Required | ||||
9 | Environmental Requirements | ||||
Operating Condition | Temperature: 0℃~40℃ Relative Humidity: 30%~85% Atmospheric Pressue: 700hPa~1060hPa | ||||
Transportation Condition | Temperature: -20℃~55℃ Relative Humidity: 20%~90% Atmospheric Pressue: 700hPa~1060hPa | ||||
Power in | 100~240V,2.7-1.1A, 50/60Hz, | ||||
10 | Qualification documents | ||||
CE certification | Required | ||||
ISO 13485 | Required |