Temperature Setting | ℃:32℃~42.0℃; ℉ :86.9℉~107.6℉ |
Temperature Accuracy | ℃:≤ ±1℃; ℉:≤ ±1.8℉ |
Increment | ℃:0.1℃; ℉:0.2℉ |
Optional Unit | ℃,℉ |
Warming Up Time | ≤2min |
Overheat Protection | 46°C,hardware and software protection |
Alarm | Overheat,low temperature,system error,door open,warming up alarm. |
Blood and IV Sets | Standard single used PVC blood or IV tubes(diameter:3.4- 4.5mm) |
Display | LED |
Heating Way | Long,Middle,short grooves |
Installation Way | IV Knob, Hanger optional |
Power Supply | Input:100-240V(AC Power)50/60HZ,0.3-0.7A Output:12V (DC Power) 2.5A |
Power Consumption | 85VA |
Operation Environment | Temp.:+5°C-+30°C; Relative Humidity:20%-80% ; Atmosphere:76.0kPa 106.0kPa |
Transportation and Storage Environment | Temp:-20°C-+55°C; Relative Humidity:10%6-93%; Atmosphere:50.0kPa-106.0kPa |
Degree of Protection against Ingress of Liquids | IPX1 |
Degree of Protection against Electric Shock | Class II, Type CF |
Working Mode | Continuous heating |
Dimension | 180×70×35mm |
Net Weight | 0.2KG;With Knob:0.4KG |