Characteristics:


1.Bigger colour touch screen with lazy operation;


2.Four-channel units design (pH/pX unit, conductive unit, DO unit) to support automatic identification, plug and play;


3.To support electrode management, up to 25pcs electrodes;


4.To support Buffer group management, automatic identification of 25 kinds of buffer solutions, such as DIN and NIST. Up to six-point pH calibration and up to 5 point conductivity calibration;


5.To support ion concentration unit switch, such as ug/L, mg/L, g/L, mol/L, mmol/L;


6.To measure 0-200ms/cm conductivity with one probe;


7.To support firm ware upgrade and contact with PC or auto sample;


8.Wide power supply from 110V to 240V


Technical data:





















































pH/pX unit



pH



Range:-2.000pH~20.000pH;


Resolution:0.001pH;


Basic error of electronic unit:±0.002pH;



pX



Range:0.000pX~14.000pX;


Resolution: 0.001pX;


Basic error of electronic unit:±0.002pX;



Ion concentration



Range:(0~19990)


Resolution:Four significant figures(Scientific notation)


Basic error of electronic unit:±0.3%



Potential



Range:-2000.00mV~2000.00mV;


Resolution:0.01mV;


Basic error of electronic unit:±0.03%FS;



Conductivityunit



Conductivity



Range:0.000μS/cm~3000mS/cm


Resolution:Varies with each grade


Basic error of electronic unit: ±0.5%(FS);



Resistivity



Range:5.00Ω.cm~100.0MΩ.cm


Resolution:Varies with each grade


Basic error of electronic unit: ±0.5%(FS);



TDS



Range:0.000 mg/L~1000g/L


Resolution:Varies with each grade


Basic error of electronic unit: ±0.5%(FS);



Salinity



Range:(0.00~8.00)%


Resolution:0.01%


Basic error of electronic unit: ±0.1%



DO unit



DO Concentration



Range:(0.00~90.00)mg/L;


Resolution:0.01mg/L ;


Basic error of electronic unit:±0.10mg/L;



DO Saturation



Range:(0.0~600.0)%


Resolution:0.1%


Basic error of electronic unit:±2.0%



Temperature



Range:(-5.0~130.0)


Resolution:0.1


Basic error of electronic unit:±0.1